The atrioventricular delay of cardiac resynchronization can be optimized hemodynamically during exercise and predicted from resting measurements

被引:25
作者
Whinnett, Zachary I. [1 ]
Briscoe, Cathy
Davies, Justin E. R.
Willson, Keith
Manisty, Charlotte H.
Davies, Wyn
Peters, Nicholas S.
Kanagaratnam, Prapa
Hughes, Alun D.
Mayet, Jamit
Francis, Darrel P.
机构
[1] St Marys Hosp, Int Ctr Circulatory Hlth, London W2 1LA, England
基金
英国惠康基金;
关键词
atrioventricular delay optimization; exercise; optimization of cardiac resynchronization therapy; biventricular pacing;
D O I
10.1016/j.hrthm.2007.11.019
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Atrioventricular (AV) optimization of cardiac resynchronization therapy (CRT) is typically calculated at rest. However, patients often become symptomatic during exercise. OBJECTIVE In this study, we use acute noninvasive hemodynamics to optimize the AV delay of CRT during exercise and investigate whether this exercise optimum can be predicted from a three-phase resting model. METHODS In 20 patients with CRT, we adjusted the sensed AV delay while the patient exercised on a treadmill up to a heart rate of 100 bpm to identify the hemodynamically optimal value. Separately, at rest, by pacing with three different configurations and calculating the sensed-paced difference, we calculated an "expected" value for the exercise optimum. RESULTS It was possible to perform AV delay optimization while a patient exercised. The resting three-phase model correlated well with the actual exercise optimal AV delay (r = 0.85, mean difference +/- standard deviation [SD] = 3.7 +/- 17 ms). Simply using measurements made at rest during atrial-sensed pacing showed a poorer correlation with exercise (r = 0.64, mean difference +/- SD = 2.2 +/- 24 ms). The three-phase resting model allows improved exercise hemodynamics to be achieved. Programming according to the three-phase resting model yields an exercise blood pressure of only 0.5 mmHg (+/- 1.4 mmHg; P = NS) Less than the true exercise optimum, whereas programming the resting sensed optimum yields an exercise blood pressure of 1.4 mmHg ( 2.2 mmHg, P =.02) less than the true optimum. CONCLUSIONS Using acute noninvasive hemodynamics and a protocol of alternations, it is possible to optimize the AV delay of cardiac resynchronization devices even while a patient exercises. In clinical practice, the exercise optimum AV delay could be determined from three phases of resting measurements, without performing exercise.
引用
收藏
页码:378 / 386
页数:9
相关论文
共 26 条
[11]   Fifteen years experience with finger arterial pressure monitoring: assessment of the technology [J].
Imholz, BPM ;
Wieling, W ;
van Montfrans, GA ;
Wesseling, KH .
CARDIOVASCULAR RESEARCH, 1998, 38 (03) :605-616
[12]   THE HEMODYNAMIC BENEFIT OF DIFFERENTIAL ATRIOVENTRICULAR DELAY INTERVALS FOR SENSED AND PACED ATRIAL EVENTS DURING PHYSIOLOGIC PACING [J].
JANOSIK, DL ;
PEARSON, AC ;
BUCKINGHAM, TA ;
LABOVITZ, AJ ;
REDD, RM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (02) :499-507
[13]   Finger arterial versus intrabrachial pressure and continuous cardiac output during head-up tilt testing in healthy subjects [J].
Jellema, WT ;
Imholz, BPM ;
vanGoudoever, J ;
Wesseling, KH ;
vanLieshout, JJ .
CLINICAL SCIENCE, 1996, 91 (02) :193-200
[14]  
Penaz J., 1973, INT C MED BIOL ENG D, P104
[15]   Simultaneous vs. sequential biventricular pacing in dilated cardiomyopathy:: an acute hemodynamic study [J].
Perego, GB ;
Chianca, R ;
Facchini, M ;
Frattola, A ;
Balla, E ;
Zucchi, S ;
Cavaglià, S ;
Vicini, I ;
Negretto, M ;
Osculati, G .
EUROPEAN JOURNAL OF HEART FAILURE, 2003, 5 (03) :305-313
[16]  
Petersen MEV, 1995, EUR HEART J, V16, P1647, DOI 10.1093/oxfordjournals.eurheartj.a060791
[17]   Rate-dependent AV delay optimization in cardiac resynchronization therapy [J].
Scharf, C ;
Li, P ;
Muntwyler, J ;
Chugh, A ;
Oral, H ;
Pelosi, F ;
Morady, F ;
Armstrong, WF .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04) :279-284
[18]   Intraventricular conduction delay: a prognostic marker in chronic heart failure [J].
Shamim, W ;
Francis, DP ;
Yousufuddin, M ;
Varney, S ;
Pieopli, MF ;
Anker, SD ;
Coats, AJS .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1999, 70 (02) :171-178
[19]  
Smith N T, 1985, J Clin Monit, V1, P17, DOI 10.1007/BF02832685
[20]   Effect of optimizing the VV interval on left ventricular contractility in cardiac resynchronization therapy [J].
van Gelder, BM ;
Bracke, FA ;
Meijer, A ;
Lakerveld, LJM ;
Pijls, NHJ .
AMERICAN JOURNAL OF CARDIOLOGY, 2004, 93 (12) :1500-1503