Effective sample size in diffuse reflectance near-IR spectrometry

被引:67
作者
Berntsson, O [1 ]
Burger, T
Folestad, S
Danielsson, LG
Kuhn, J
Fricke, J
机构
[1] Royal Inst Technol, Div Analyt Chem, Dept Chem, SE-10044 Stockholm, Sweden
[2] Univ Wurzburg, Inst Phys, D-97074 Wurzburg, Germany
[3] Astra Hassle AB, Pharmaceut R&D, Analyt Res, SE-43183 Molndal, Sweden
[4] ZAE Bayern, Bayer Zentrum Angew Energieforsch, D-97074 Wurzburg, Germany
关键词
D O I
10.1021/ac980652u
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Two independent methods for determination of the effectively sampled mass per unit area are presented and compared. The first method combines directional-hemispherical transmittance and reflectance measurements. A three-flux approximation of the equation of radiative transfer is used, to separately determine the specific absorption and scattering coefficients of the powder material, which subsequently are used to determine the effective sample size. The second method uses a number of diffuse reflectance measurements on layers of controlled powder thickness in an empirical approach. The two methods are shown to agree well and thus confirm each other. From the determination of the effective sample size at each measured wavelength in the visible-NIR region for two different model powder materials, large differences was found, both between the two analyzed powders and between different wavelengths. As an example, the effective sample size ranges between 15 and 70 mg/cm(2) for microcrystalline cellulose and between 70 and 300 mg/cm(2) for film-coated pellets. However, the contribution to the spectral information obtained from a certain layer decreases rapidly with increasing distance from the powder surface. With both methods, the extent of contribution from various depths of a powder sample to the visible-NIR diffuse reflection signal is characterized. This information is valuable for validation of analytical applications of diffuse reflectance visible-NIR spectrometry.
引用
收藏
页码:617 / 623
页数:7
相关论文
共 37 条
[11]  
CIURCZAK EW, 1992, SPECTROSCOPY, V7, P12
[12]  
CORTI P, 1993, FARMACO, V48, P3
[13]   Automated system for the on-line monitoring of powder blending processes using near-infrared spectroscopy .1. System development and control [J].
Hailey, PA ;
Doherty, P ;
Tapsell, P ;
Oliver, T ;
Aldridge, PK .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1996, 14 (05) :551-559
[14]  
HERSEY JA, 1979, INT J PHAR TECH PROD, V1, P6
[15]  
KAGANER MG, 1969, OPT SPECTROSC-USSR, V26, P240
[16]  
Kaviany M., 1991, PRINCIPLES HEAT TRAN, DOI DOI 10.1007/978-1-4684-0412-8
[17]  
KEMENY GJ, 1992, HDB NEAR INFRARED AN, pCH5
[18]   NEAR-INFRARED SPECTROSCOPY - APPLICATIONS IN THE ANALYSIS OF TABLETS AND SOLID PHARMACEUTICAL DOSAGE FORMS [J].
KIRSCH, JD ;
DRENNEN, JK .
APPLIED SPECTROSCOPY REVIEWS, 1995, 30 (03) :139-174
[19]  
Kortum G., 1969, REFLEXIONSSPEKTROSKO
[20]   INFRARED-OPTICAL TRANSMISSION AND REFLECTION MEASUREMENTS ON LOOSE POWDERS [J].
KUHN, J ;
KORDER, S ;
ARDUINISCHUSTER, MC ;
CAPS, R ;
FRICKE, J .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1993, 64 (09) :2523-2530