A liquid chromatography assay for a quantification of doripenem, ertapenem, imipenem, meropenem concentrations in human plasma: Application to a clinical pharmacokinetic study

被引:67
作者
Dailly, E. [1 ,2 ]
Bouquie, R. [1 ,2 ]
Deslandes, G. [1 ]
Jolliet, P. [1 ,2 ]
Le Floch, R. [3 ]
机构
[1] CHU Nantes, Dept Clin Pharmacol, F-44035 Nantes 01, France
[2] Univ Nantes, Fac Med Pharm, EA Biostat Rech Clin & Mesures Subject Sante 4275, F-44035 Nantes, France
[3] CHU Nantes, Burn Care Dept, F-44035 Nantes 01, France
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2011年 / 879卷 / 15-16期
关键词
Doripenem; Ertapenem; Meropenem; Imipenem; Ultra high efficiency chromatography; ULTRAVIOLET DETECTION; BRONCHOALVEOLAR LAVAGE; MOUSE PLASMA; HUMAN SERUM; PERFORMANCE; ANTIBIOTICS; FLUID; HPLC; CARBAPENEMS; CEFEPIME;
D O I
10.1016/j.jchromb.2011.03.038
中图分类号
Q5 [生物化学];
学科分类号
070307 [化学生物学];
摘要
A simple chromatographic assay based on ultra high performance liquid chromatography with ultraviolet detection at 295 nm is proposed to determinate simultaneously human plasma concentrations of imipenem, doripenem, meropenem and ertapenem. After deproteinization by acetonitrile, carbapenems are separated on a PentaFluoroPhenyl column with a binary gradient elution. This method is specific, accurate, precise (the intra-day and inter-day imprecision and inaccuracy are lower than 15%), sensitive (the limit of quantitation is equal to 0.50 mg/L for imipenem, doripenem, ertapenem, meropenem) and not time consuming (run time = 7 min). An application of this method to measure ertapenem plasma concentrations in burn patients is presented. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:1137 / 1142
页数:6
相关论文
共 30 条
[1]
Liquid chromatographic method for the simultaneous determination of imipenem and sulbactam in mouse plasma [J].
Aparicio, Irene ;
Bello, Miguel Angel ;
Callejon, Manuel ;
Jimenez, Juan Carlos .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2006, 44 (09) :548-551
[2]
Determination of meropenem in serum by high-performance liquid chromatography with column switching [J].
Bompadre, S ;
Ferrante, L ;
De Martinis, M ;
Leone, L .
JOURNAL OF CHROMATOGRAPHY A, 1998, 812 (1-2) :249-253
[3]
Ertapenem [J].
Congeni, Blaise L. .
EXPERT OPINION ON PHARMACOTHERAPY, 2010, 11 (04) :669-672
[4]
A simple and fast method for quantification of ertapenem using meropenem as internal standard in human plasma in a clinical setting [J].
D'Avolio, Antonio ;
Baietto, Lorena ;
De Rosa, Francesco Giuseppe ;
Garazzino, Silvia ;
Sciandra, Mauro ;
Siccardi, Marco ;
Bonora, Stefano ;
Di Perri, Giovanni .
THERAPEUTIC DRUG MONITORING, 2008, 30 (01) :90-94
[5]
Simultaneous determination of five β-lactam antibiotics (cefepim, ceftazidim, cefuroxim, meropenem and piperacillin) in human plasma by high-performance liquid chromatography with ultraviolet detection [J].
Denooz, Raphael ;
Charlier, Corinne .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 864 (1-2) :161-167
[6]
Rapid antibiotic drug monitoring:: Meropenem and ceftazidime determination in serum and bronchial secretions by high-performance liquid chromatography-integrated sample preparation [J].
Ehrlich, M ;
Daschner, FD ;
Kümmerer, K .
JOURNAL OF CHROMATOGRAPHY B, 2001, 751 (02) :357-363
[7]
High-performance liquid chromatographic assay for meropenem in serum [J].
Elkhaili, H ;
Niedergang, S ;
Pompei, D ;
Linger, L ;
Leveque, D ;
Jehl, F .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 1996, 686 (01) :19-26
[8]
Determination of imipenem and rifampicin in mouse plasma by high performance liquid chromatography-diode array detection [J].
Fernandez-Torres, R. ;
Bello-Lopez, M. A. ;
Callejon-Mochon, M. ;
Jimenez-Sanchez, J. C. .
ANALYTICA CHIMICA ACTA, 2008, 608 (02) :204-210
[9]
Determination of free ertapenem in plasma and bronchoalveolar lavage by high-performance liquid chromatography with ultraviolet detection [J].
Gordien, JB ;
Boselli, E ;
Fleureau, C ;
Allaouchiche, B ;
Janvier, G ;
Lalaude, O ;
Saux, MC ;
Breilh, D .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 830 (02) :218-223
[10]
Performance of columns packed with the new shell Kinetex-C18 particles in gradient elution chromatography [J].
Gritti, Fabrice ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (10) :1604-1615