Objective measurements of image quality

被引:61
作者
Browne, JE
Watson, AJ
Gibson, NM
Dudley, NJ
Elliott, AT
机构
[1] Univ Glasgow, Western Infirm, Dept Clin Phys, Glasgow G11 6NT, Lanark, Scotland
[2] Univ Glasgow, Western Infirm, Ultrasound Equipment Evaluat Project, Glasgow G11 6NT, Lanark, Scotland
[3] City Hosp Nottingham, Dept Med Phys, Nottingham, England
关键词
computerised image analysis program; tissue harmonic imaging; compound imaging; harmonic compound imaging; quality control;
D O I
10.1016/j.ultrasmedbio.2003.10.002
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Tissue harmonic imaging (THI) and compound imaging have been reported clinically to improve contrast resolution, tissue differentiation and overall image quality. However, there have been limited studies to date to quantify objectively the improvements in image quality achieved with these new imaging techniques. The aim of this study was to quantify differences in image quality that exist between conventional B-mode imaging, harmonic imaging, compound imaging and harmonic compound imaging. An ATL HDI 5000 scanner with three probes (C5-2, L74 and L12-5) was tested with two different types of test object, the Gammex-RMI model 404GS LE and the Gammex-RMI 403 GS LE. The measurement limitations associated with subjective analysis methods were not present in this study because an automated image analysis program was used to determine the image quality parameters. Therefore, subtle differences between the four imaging modes could be detected. Significant improvements in lateral resolution and slice thickness as a function of depth were found with THI Contrast resolution and anechoic target detection improved with compound imaging, and harmonic compound imaging improved lateral resolution, slice thickness as a function of depth and contrast resolution.
引用
收藏
页码:229 / 237
页数:9
相关论文
共 21 条
[1]  
ALTMAN DG, 1991, PRACTICAL STAT MED R, P293
[2]  
Averkiou MA, 1997, P IEEE ULTR S, V2, P1561
[3]   Finite amplitude distortion-based inhomogeneous pulse echo ultrasonic imaging [J].
Christopher, T .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1997, 44 (01) :125-139
[4]   Advances in ultrasound [J].
Claudon, M ;
Tranquart, F ;
Evans, DH ;
Lefèvre, F ;
Correas, JM .
EUROPEAN RADIOLOGY, 2002, 12 (01) :7-18
[5]  
Desser T. S., 2000, ULTRASOUND Q, V16, P40
[6]   Tissue harmonic imaging techniques: Physical principles and clinical applications [J].
Desser, TS ;
Jeffrey, RB .
SEMINARS IN ULTRASOUND CT AND MRI, 2001, 22 (01) :1-10
[7]   Real-time spatial compound imaging: Application to breast, vascular, and musculoskeletal ultrasound [J].
Entrekin, RR ;
Porter, BA ;
Sillesen, HH ;
Wong, AD ;
Cooperberg, PL ;
Fix, CH .
SEMINARS IN ULTRASOUND CT AND MRI, 2001, 22 (01) :50-64
[8]  
Entrekin RR, 2000, ACOUST IMAG, V25, P331
[9]   PHASING OUT SPECKLE [J].
GATENBY, JC ;
HODDINOTT, JC ;
LEEMAN, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (11) :1683-1689
[10]   A computerised quality control testing system for B-mode ultrasound [J].
Gibson, NM ;
Dudley, NJ ;
Griffith, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (12) :1697-1711