Effect of practical layered dielectric loads on SAR patterns from dual concentric conductor microstrip antennas

被引:18
作者
Rossetto, F
Stauffer, PR [1 ]
Manfrini, V
Diederich, CJ
Gentili, GB
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA 94143 USA
[2] Univ Florence, Dept Elect Engn, Florence, Italy
[3] Hewlett Packard Co, Bergamo, Italy
关键词
microwave array applicators; FDTD analysis; microstrip antennas; hyperthermia; SAR;
D O I
10.3109/02656739809018254
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Radiation patterns of 2 and 4cm square Dual Concentric Conductor (DCC) microstrip antennas were studied theoretically with Finite Difference Time Domain (FDTD) analysis and compared with experimental measurements of power deposition (SAR) in layered lossy dielectric loads. Single and array configurations were investigated with 915 MHz excitation applied across either one, two or four sides, or four corners of the square apertures. FDTD simulations were carried out for realistic models of a muscle tissue load coupled to the DCC antennas with a 5 mm thick bolus of either distilled water or low loss Silicone Oil. This study characterizes the effect on SAR of adding three additional thin dielectric layers which are necessary for clinical use of the applicator. These layers consist of a 0.1mm thick dielectric coating on the array surface to provide electrical isolation of DCC apertures, and 0.15mm thick plastic layers above and below the bolus to contain the liquid. Experimental measurements of SAR in a plane 1 cm deep in muscle phantom agree well with theoretical FDTD simulations in the multi-layered tissue models. These studies reveal significant changes in SAR for applicator configurations involving low dielectric constant (epsilon(r)) layers on either side of a high epsilon(r) water bolus layer. Prominent changes include a broadening and centring of the SAR under each aperture as well as increased SAR penetration in muscle. No significant differences are noted between the simple and complete load configurations for the low epsilon(r) Silicone Oil bolus. Both theoretical and measured data demonstrate relatively uniform SAR distributions with > 50% of maximum SAR extending to the perimeter of single and multi-aperture array configurations of DCC applicators when using a thin 5 mm water or Silicone Oil bolus.
引用
收藏
页码:553 / 571
页数:19
相关论文
共 24 条
[1]   EVALUATION OF CAPTIVE BOLUS APPLICATORS [J].
CHOU, CK ;
MCDOUGALL, JA ;
CHAN, KW ;
LUK, KH .
MEDICAL PHYSICS, 1990, 17 (04) :705-709
[2]   PRECLINICAL EVALUATION OF A MICROWAVE PLANAR ARRAY APPLICATOR FOR SUPERFICIAL HYPERTHERMIA [J].
DIEDERICH, CJ ;
STAUFFER, PR .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1993, 9 (02) :227-246
[3]   CURRENT SHEET APPLICATOR ARRAYS FOR SUPERFICIAL HYPERTHERMIA OF CHESTWALL LESIONS [J].
GOPAL, MK ;
HAND, JW ;
LUMORI, MLD ;
ALKHAIRI, S ;
PAULSEN, KD ;
CETAS, TC .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (02) :227-240
[5]  
GUY AW, 1968, THERMAL PROBLEMS BIO, P26
[6]   SIMULATED BIOLOGICAL-MATERIALS FOR ELECTROMAGNETIC-RADIATION ABSORPTION STUDIES [J].
HARTSGROVE, G ;
KRASZEWSKI, A ;
SUROWIEC, A .
BIOELECTROMAGNETICS, 1987, 8 (01) :29-36
[7]   COMPARATIVE HEATING-PATTERN STUDY OF DIRECT-CONTACT APPLICATORS IN MICROWAVE DIATHERMY [J].
KANTOR, G ;
CETAS, TC .
RADIO SCIENCE, 1977, 12 (06) :111-120
[8]   Efficacy of adjuvant hyperthermia in the treatment of superficial recurrent breast cancer: Confirmation and future directions [J].
Kapp, DS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (05) :1117-1121
[9]  
MANFRINI V, 1995, P WESC 95 SAN FRANC, P662
[10]   Hyperthermia as an adjuvant to radiation therapy of recurrent or metastatic malignant melanoma. A multicentre randomized trial by the European Society for Hyperthermic Oncology [J].
Overgaard, J ;
Gonzalez, DG ;
Hulshof, MCCH ;
Arcangeli, G ;
Dahl, O ;
Mella, O ;
Bentzen, SM .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1996, 12 (01) :3-20