We have developed a fast calibration method for computing the position and orientation of 2-D ultrasound (US) images in 3-D space where a position sensor is mounted on the US probe. This calibration is required in the fields of 3-D ultrasound and registration of ultrasound with other imaging modalities. Most of the existing calibration methods require a complex and tedious experimental procedure. Our method is simple and it is based on a custom-built phantom. Thirty N-fiducials (markers in the shape of the letter "N") embedded in the phantom provide the basis for our calibration procedure. We calibrated a 3.5-MHz sector phased-array probe with a magnetic position sensor, and we studied the accuracy and precision of our method. A typical calibration procedure requires approximately 2 min. We conclude that we can achieve accurate and precise calibration using a single US image, provided that a large number (approximately ten) of N-fiducials are captured within the US image, enabling a representative sampling of the imaging plane.
引用
收藏
页码:1219 / 1229
页数:11
相关论文
共 15 条
[1]   LEAST-SQUARES FITTING OF 2 3-D POINT SETS [J].
ARUN, KS ;
HUANG, TS ;
BLOSTEIN, SD .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (05) :699-700
[2]   STEREOTACTIC HEAD FRAME FOR USE WITH CT BODY SCANNERS [J].
BROWN, RA .
INVESTIGATIVE RADIOLOGY, 1979, 14 (04) :300-304
[3]   Intraoperative US in interactive image-guided neurosurgery [J].
Comeau, RM ;
Fenster, A ;
Peters, TM .
RADIOGRAPHICS, 1998, 18 (04) :1019-1027
[4]   3D ULTRASONIC IMAGE FEATURE LOCALIZATION BASED ON MAGNETIC SCANHEAD TRACKING - IN-VITRO CALIBRATION AND VALIDATION [J].
DETMER, PR ;
BASHEIN, G ;
HODGES, T ;
BEACH, KW ;
FILER, EP ;
BURNS, DH ;
STRANDNESS, DE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (09) :923-936
[5]  
Edwards WS, 1998, INT J IMAG SYST TECH, V9, P442, DOI 10.1002/(SICI)1098-1098(1998)9:6<442::AID-IMA6>3.0.CO
[6]  
2-7
[7]   Development of a frameless and armless stereotactic neuronavigation system with ultrasonographic registration [J].
Hata, N ;
Dohi, T ;
Iseki, H ;
Takakura, K .
NEUROSURGERY, 1997, 41 (03) :608-613
[8]   Performance of a miniature magnetic position sensor for three-dimensional ultrasound imaging [J].
Leotta, DF ;
Detmer, PR ;
Martin, RW .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (04) :597-609
[9]   Registration of head volume images using implantable fiducial markers [J].
Maurer, CR ;
Fitzpatrick, JM ;
Wang, MY ;
Galloway, RL ;
Maciunas, RJ ;
Allen, GS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (04) :447-462
[10]  
Pagoulatos N, 1999, IEEE Trans Inf Technol Biomed, V3, P278, DOI 10.1109/4233.809172