Impact of transvenous lead position on active-can ICD defibrillation: A computer simulation study

被引:22
作者
Aguel, F
Eason, JC
Trayanova, NA
Siekas, G
Fishler, MG
机构
[1] Tulane Univ, Dept Biomed Engn, New Orleans, LA 70118 USA
[2] Univ Memphis, Dept Biomed Engn, Memphis, TN 38152 USA
[3] St Jude Med CRMD, Sunnyvale, CA USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1999年 / 22卷 / 01期
关键词
transvenous defibrillation thresholds; torso; FEM; ICD;
D O I
10.1111/j.1540-8159.1999.tb00324.x
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Optimizing lead placement in transvenous defibrillation remains central to the clinical aspects of the defibrillation procedure. Studies involving superior vena cava (SVC) return electrodes have found that left ventricular (LV) leads or septal positioning of the right ventricular (RV) lead minimizes the voltage defibrillation threshold (VDFT) in endocardial lead-->SVC defibrillation systems. However, similar studies have not been conducted for active-can configurations. The goal of this study was to determine the optimal lead position to minimize the VDFT for systems incorporating an active can. This study used a high resolution finite element model of a human torso that includes the fiber architecture of the ventricular myocardium to find the role of lead positioning in a transvenous LEAD-->can defibrillation electrode system. It was found that, among single lead systems, posterior positioning of leads in the right ventricle lowers VDFTs appreciably. Furthermore, a septal location of leads resulted in lower VDFTs than free-wall positioning. Increasing the number of leads, and thus the effective lead surface area in the right ventricle also resulted in lower VDFTs. However, the lead configuration that resulted in the lowest VDFTs is a combination of a mid-cavity right ventricle lead and a mid-cavity left ventricle lead. The addition of a left ventricular lead resulted in a reduction in the size of the low gradient regions and a change of its location from the left ventricular free wall to the septal wall.
引用
收藏
页码:158 / 164
页数:7
相关论文
共 22 条
[1]
A 3-DIMENSIONAL FINITE-ELEMENT MODEL OF HUMAN TRANSTHORACIC DEFIBRILLATION - PADDLE PLACEMENT AND SIZE [J].
CAMACHO, MA ;
LEHR, JL ;
EISENBERG, SR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (06) :572-578
[2]
DIRECTIONAL DIFFERENCES OF IMPULSE SPREAD IN TRABECULAR MUSCLE FROM MAMMALIAN HEART [J].
CLERC, L .
JOURNAL OF PHYSIOLOGY-LONDON, 1976, 255 (02) :335-346
[3]
EASON J, 1998, IN PRESS ANN BIOMED
[4]
STIMULUS-INDUCED CRITICAL-POINT - MECHANISM FOR ELECTRICAL INITIATION OF REENTRY IN NORMAL CANINE MYOCARDIUM [J].
FRAZIER, DW ;
WOLF, PD ;
WHARTON, JM ;
TANG, ASL ;
SMITH, WM ;
IDEKER, RE .
JOURNAL OF CLINICAL INVESTIGATION, 1989, 83 (03) :1039-1052
[5]
SPECIFIC RESISTANCE OF BIOLOGICAL MATERIAL-A COMPENDUM OF DATA FOR BIOMEDICAL ENGINEER AND PHYSIOLOGIST [J].
GEDDES, LA ;
BAKER, LE .
MEDICAL & BIOLOGICAL ENGINEERING, 1967, 5 (03) :271-&
[6]
Electrical defibrillation optimization: An automated, iterative parallel finite-element approach [J].
Hutchinson, SA ;
Ng, KT ;
Shadid, JN ;
Nadeem, A .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1997, 44 (04) :278-289
[7]
CURRENT CONCEPTS FOR SELECTING THE LOCATION, SIZE AND SHAPE OF DEFIBRILLATION ELECTRODES [J].
IDEKER, RE ;
WOLF, PD ;
ALFERNESS, C ;
KRASSOWSKA, W ;
SMITH, WM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (02) :227-&
[8]
PREDICTING CARDIOTHORACIC VOLTAGES DURING HIGH-ENERGY SHOCKS - METHODOLOGY AND COMPARISON OF EXPERIMENTAL TO FINITE-ELEMENT MODEL DATA [J].
JORGENSON, DB ;
SCHIMPF, PH ;
SHEN, I ;
JOHNSON, G ;
BARDY, GH ;
HAYNOR, DR ;
KIM, YM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (06) :559-571
[9]
COMPUTATIONAL STUDIES OF TRANSTHORACIC AND TRANSVENOUS DEFIBRILLATION IN A DETAILED 3-D HUMAN THORAX MODEL [J].
JORGENSON, DB ;
HAYNOR, DR ;
BARDY, GH ;
KIM, YM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (02) :172-184
[10]
FINITE-ELEMENT MODELS OF THORACIC CONDUCTIVE ANATOMY - SENSITIVITY TO CHANGES IN INHOMOGENEITY AND ANISOTROPY [J].
KARLON, WJ ;
LEHR, JL ;
EISENBERG, SR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (11) :1010-1017