Design of experiment assisted concurrent enantioseparation of bupropion and hydroxybupropion by high-performance thin-layer chromatography

被引:9
作者
Bhatt, Nejal M. [1 ]
Chavada, Vijay D. [1 ]
Sanyal, Mallika [2 ]
Shrivastav, Pranav S. [1 ]
机构
[1] Gujarat Univ, Sch Sci, Dept Chem, Ahmadabad, Gujarat, India
[2] St Xaviers Sci Coll, Dept Chem, Ahmadabad, Gujarat, India
关键词
Box Behnken model; bupropion; hydroxybupropion; chiral separation; design of experiment; high-performance thin-layer chromatography; TANDEM MASS-SPECTROMETRY; LIQUID-CHROMATOGRAPHY; HUMAN PLASMA; MAJOR METABOLITES; ASSAY;
D O I
10.1002/chir.22673
中图分类号
R914 [药物化学];
学科分类号
100705 [微生物与生化药学];
摘要
A simple and efficient high-performance thin-layer chromatographic method was developed for chiral separation of rac-bupropion (BUP) and its active metabolite rac-hydroxybupropion (HBUP). Design of experiment (DoE)-based optimization was adopted instead of a conventional trial-and-error approach. The Box-Behnken design surface response model was used and the operating variables were optimized based on 17 trials design. The optimized method involved impregnation of chiral reagent, L(+)-tartaric acid, in the stationary phase with simultaneous addition in the mobile phase, which consisted of acetonitrile : methanol : dichloromethane : 0.50% L-tartaric acid (6.75: 1.0: 1.0: 0.25, v/v/v/v). Under the optimized conditions, the resolution factor between the enantiomers of BUP and HBUP was 6.30 and 9.26, respectively. The limit of detection and limit of quantitation for (R)-BUP, (S)-BUP, (R, R)-HBUP, and (S, S)-HBUP were 9.23 and 30.78 ng spot(-1), 10.32 and 34.40 ng spot(-1), 12.19 and 40.65 ng spot(-1), and 14.26 and 47.53 ng spot(-1), respectively. The interaction of L-tartaric acid with analytes and their retention behavior was thermodynamically investigated using van't Hoff's plots. The developed method was validated as per the International Conference on Harmonization guidelines. Finally, the method was successfully applied to resolve and quantify the enantiomeric content from marketed tablets as well as spiked plasma samples.
引用
收藏
页码:80 / 88
页数:9
相关论文
共 19 条
[1]
Batra S, 2015, BIOMED CHROMATOGR, V30, P370
[2]
High-performance liquid chromatographic enantioseparation of (R,S)-fluoxetine using Marfey's reagent and (S)-N-(4-nitrophenoxycarbonyl) phenylalanine methoxyethyl ester as chiral derivatizing reagents along with direct thin-layer chromatographic resolution and isolation of enantiomers using L-tartaric acid as mobile phase additive [J].
Bhushan, Ravi ;
Agarwal, Charu .
BIOMEDICAL CHROMATOGRAPHY, 2010, 24 (11) :1152-1158
[3]
High-throughput liquid chromatography-tandem mass spectrometry determination of bupropion and its metabolites in human, mouse and rat plasma using a monolithic column [J].
Borges, V ;
Yang, E ;
Dunn, J ;
Henion, J .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 804 (02) :277-287
[4]
Stereoselective analysis of bupropion and hydroxybupropion in human plasma and urine by LC/MS/MS [J].
Coles, Rebecka ;
Kharasch, Evan D. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 857 (01) :67-75
[5]
DETERMINATION OF BUPROPION AND ITS MAJOR BASIC METABOLITES IN PLASMA BY LIQUID-CHROMATOGRAPHY WITH DUAL-WAVELENGTH ULTRAVIOLET DETECTION [J].
COOPER, TB ;
SUCKOW, RF ;
GLASSMAN, A .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1984, 73 (08) :1104-1107
[6]
Croarkin C., NIST SEMATECH E HDB
[7]
THE FDA PERSPECTIVE ON THE DEVELOPMENT OF STEREOISOMERS [J].
DECAMP, WH .
CHIRALITY, 1989, 1 (01) :2-6
[8]
Ultra-Performance Liquid Chromatography-Tandem Mass Spectrometry Method for the Determination of Bupropion and its Main Metabolites in Human Whole Blood [J].
Denooz, Raphael ;
Mercerolle, Magali ;
Lachatre, Gerard ;
Charlier, Corinne .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2010, 34 (05) :280-286
[9]
International Conference on Harmonization, 2005, VAL AN PROC TEXT MET
[10]
HPLC assay for bupropion and its major metabolites in human plasma [J].
Loboz, KK ;
Gross, AS ;
Ray, J ;
McLachlan, AJ .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 823 (02) :115-121