A bienzyme channeling glucose sensor with a wide concentration range based on co-entrapment of enzymes in SBA-15 mesopores

被引:97
作者
Dai, Zhihui [1 ]
Bao, Hanchun
Yang, Xiaodi
Ju, Huangxian [2 ]
机构
[1] Nanjing Normal Univ, Coll Chem & Environm Sci, Jiangsu Key Lab Bionfunct Mat, Nanjing 210097, Peoples R China
[2] Nanjing Univ, Dept Chem, MOE Key lab Analyt Chem Life Sci, Nanjing 210093, Peoples R China
基金
中国国家自然科学基金;
关键词
glucose oxidase; horseradish peroxidase; enzyme-channeling sensor; SBA-15; glucose;
D O I
10.1016/j.bios.2007.10.015
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A novel bienzyme-channeling sensor was constructed by entrapping glucose oxidase (GOD) and horseradish peroxidase (HRP) in the mesopores of well-ordered hexagonal mesoporous silica structures (SBA-15). The SBA-15 mesoporous materials accelerated the electron transfer between the entrapped HRP and electrode. Both HRP and GOD retained their catalytic activities in the bienzyme-entrapped SBA-15 film. In presence of glucose the enzymatic reaction of GOD-glucose-dissolved oxygen system generated hydrogen peroxide in the bienzyme-entrapped mesopores, which was immediately reduced at -0.40 V by an electrocatalytic reaction with the HRP entrapped in the same mesopore to lead to a sensitive and fast amperometric response. Thus the bienzyme channeling could be used for the detection of glucose with excellent performance without the addition of any mediator. Optimization of the experimental parameters was performed with regard to pH, operating potential and temperature. The detection limit was down to 2.7 x 10(-7) M with a very wide linear range from 3.0 x 10(-6) to 3.4 x 10(-2) M. The constructed bienzyme channeling provided a strategy for amperometric detection of oxidase substrates by co-entrapping the corresponding oxidase and HRP in the mesoporous materials. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:1070 / 1076
页数:7
相关论文
共 45 条
  • [1] THE DETERMINATION OF PORE VOLUME AND AREA DISTRIBUTIONS IN POROUS SUBSTANCES .1. COMPUTATIONS FROM NITROGEN ISOTHERMS
    BARRETT, EP
    JOYNER, LG
    HALENDA, PP
    [J]. JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1951, 73 (01) : 373 - 380
  • [2] A wireless, remote query glucose biosensor based on a pH-sensitive polymer
    Cai, QY
    Zeng, KF
    Ruan, CM
    Desai, TA
    Grimes, CA
    [J]. ANALYTICAL CHEMISTRY, 2004, 76 (14) : 4038 - 4043
  • [3] Characterization of the direct electron transfer and bioelectrocatalysis of horseradish peroxidase in DNA film at pyrolytic graphite electrode
    Chen, XH
    Ruan, CM
    Kong, JL
    Deng, JQ
    [J]. ANALYTICA CHIMICA ACTA, 2000, 412 (1-2) : 89 - 98
  • [4] ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY
    CLARK, LC
    LYONS, C
    [J]. ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1962, 102 (01) : 29 - &
  • [5] Cosnier S, 2000, ELECTROANAL, V12, P356, DOI 10.1002/(SICI)1521-4109(20000301)12:5<356::AID-ELAN356>3.3.CO
  • [6] 2-Q
  • [7] Specific determination of As(V) by an acid phosphatase-polyphenol oxidase biosensor
    Cosnier, Serge
    Mousty, Christine
    Cui, Xiaoqiang
    Yang, Xiurong
    Dong, Shaojun
    [J]. ANALYTICAL CHEMISTRY, 2006, 78 (14) : 4985 - 4989
  • [8] Amperometric, screen-printed, glucose biosensor for analysis of human plasma samples using a biocomposite water-based carbon ink incorporating glucose oxidase
    Crouch, E
    Cowell, DC
    Hoskins, S
    Pittson, RW
    Hart, JP
    [J]. ANALYTICAL BIOCHEMISTRY, 2005, 347 (01) : 17 - 23
  • [9] Biosensors in clinical chemistry
    D'Orazio, P
    [J]. CLINICA CHIMICA ACTA, 2003, 334 (1-2) : 41 - 69
  • [10] Direct electrochemistry of glucose oxidase immobilized on a hexagonal mesoporous silica-MCM-41 matrix
    Dai, Z. H.
    Ni, J.
    Huang, X. H.
    Lu, G. F.
    Bao, J. C.
    [J]. BIOELECTROCHEMISTRY, 2007, 70 (02) : 250 - 256