Effect of irradiation type (LED or QTH) on photo-activated composite shrinkage strain kinetics, temperature rise, and hardness

被引:125
作者
Hofmann, N [1 ]
Hugo, B [1 ]
Klaiber, B [1 ]
机构
[1] Univ Wurzburg, Dept Operat Dent & Periodontol, D-97070 Wurzburg, Germany
关键词
composite resins; radiation effects; kinetics; heat; hardness tests;
D O I
10.1034/j.1600-0722.2002.21359.x
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
This study compares commercially available light-emitting diode ( LED) lights with a quartz tungsten halogen (QTH) unit for photo-activating resin-based composites (RBC). Shrinkage strain kinetics and temperature within the RBC were measured simultaneously using the 'deflecting disc technique' and a thermocouple. Surface hardness ( Knoop) at the bottom of 1.5-mm thick RBC specimens was measured 24 h after irradiation to indicate degree of cure. Irradiation was performed for 40 s using either the continuous or the ramp-curing mode of a QTH and a LED light (800 mW cm(-2) and 320 mW cm(-2), respectively) or the continuous mode of a lower intensity LED light ( 160 mW cm(-2)). For Herculite XRV and Filtek Z250 ( both containing only camphoroquinone as a photo-initiator) the QTH and the stronger LED light produced similar hardness, while in the case of Definite ( containing an additional photo-activator absorbing at lower wavelength) lower hardness was observed after LED irradiation. The temperature rise during polymerization and heating from radiation were lower with LED compared to QTH curing. The fastest increase of polymerization contraction was observed after QTH continuous irradiation, followed by the stronger and the weaker LED light in the continuous mode. Ramp curing decreased contraction speed even more. Shrinkage strain after 60 min was greater following QTH irradiation compared with both LED units ( Herculite, Definite) or with the weaker LED light (Z250).
引用
收藏
页码:471 / 479
页数:9
相关论文
共 44 条
[1]
Effect of photopolymerization variables on composite hardness [J].
Abate, PF ;
Zahra, VN ;
Macchi, RL .
JOURNAL OF PROSTHETIC DENTISTRY, 2001, 86 (06) :632-635
[2]
Blankenau R J, 1991, Am J Dent, V4, P40
[3]
Bouschlicher M R, 2000, J Esthet Dent, V12, P328, DOI 10.1111/j.1708-8240.2000.tb00242.x
[4]
Bouschlicher M R, 2000, J Esthet Dent, V12, P23, DOI 10.1111/j.1708-8240.2000.tb00195.x
[5]
Polymerization contraction stress in light-cured packable composite resins [J].
Chen, HY ;
Manhart, J ;
Hickel, R ;
Kunzelmann, KH .
DENTAL MATERIALS, 2001, 17 (03) :253-259
[6]
SPECTRAL DISTRIBUTIONS OF DENTAL PHOTO-POLYMERIZATION SOURCES [J].
COOK, WD .
JOURNAL OF DENTAL RESEARCH, 1982, 61 (12) :1436-1438
[7]
RELAXATION OF POLYMERIZATION CONTRACTION STRESSES BY FLOW IN DENTAL COMPOSITES [J].
DAVIDSON, CL ;
DEGEE, AJ .
JOURNAL OF DENTAL RESEARCH, 1984, 63 (02) :146-148
[8]
The effect of curing light variations on bulk curing and wall-to-wall quality of two types and various shades of resin composites [J].
Davidson-Kaban, SS ;
Davidson, CL ;
Feilzer, AJ ;
de Gee, AJ ;
Erdilek, N .
DENTAL MATERIALS, 1997, 13 (5-6) :344-352
[9]
INFLUENCE OF LIGHT-INTENSITY ON POLYMERIZATION SHRINKAGE AND INTEGRITY OF RESTORATION-CAVITY INTERFACE [J].
FEILZER, AJ ;
DOOREN, LH ;
DEGEE, AJ ;
DAVIDSON, CL .
EUROPEAN JOURNAL OF ORAL SCIENCES, 1995, 103 (05) :322-326
[10]
Ferracane J L, 1985, Dent Mater, V1, P11, DOI 10.1016/S0109-5641(85)80058-0