Caval incorporation of the LGM Vena Tech filter: An experimental study

被引:12
作者
Crochet, D
BachLijour, B
Grossetete, R
Raffin, T
Nguyen, JM
Hurtrel, M
Daculsi, G
LeNihouannen, JC
机构
[1] ECOLE NATL VET,DEPT SURG,F-44087 NANTES 03,FRANCE
[2] UNIV NANTES,RES LAB CALCIFIED TISSUES & BIOMAT,F-44002 NANTES,FRANCE
关键词
embolism; pulmonary; interventional procedures; experimental studies; venae cavae; filters;
D O I
10.1016/S1051-0443(97)70582-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To analyze LGM Vena Tech filter incorporation and the rapidity of the process. MATERIALS AND METHODS: A filter was inserted into the infrarenal inferior vena cava (IVC) Of 15 ewes assigned to one of three groups depending on the length of follow-up (2, 4, or 8 weeks), Radiologic data concerning IVC diameter and filter patency, stability, and incorporation were obtained before and after insertion and before euthanasia. Histopathologic analysis concerned wall thickness and smooth muscle cell area (SMCA) at three levels of the filter and at one point outside the filter. RESULTS:. All filters remained patent during follow-up, Incorporation of struts was dependent on time (P = .006), level of the filter (P = .0001), and strut surface (P < .0001), Neointimal thickness increased during follow-up (P = .0002), being more marked in the midportion of the filter (P = .0037), Adventitial thinning was observed (P = .0001), corresponding to a significant decrease in SMCA (P < .0001) above the struts as a function of the length of the follow-up period (P = .0021). CONCLUSIONS: The LGM Vena Tech filter was well tolerated and is suitable for incorporation into the IVC wall of normal animals without risk of any deleterious reactions due to biological incompatibility.
引用
收藏
页码:419 / 425
页数:7
相关论文
共 23 条
[1]   FLEXIBLE TANTALUM STENTS IMPLANTED IN AORTAS AND ILIAC ARTERIES - EFFECTS IN NORMAL CANINES [J].
BARTH, KH ;
VIRMANI, R ;
STRECKER, EP ;
SAVIN, MA ;
LINDISCH, D ;
MATSUMOTO, AH ;
TEITELBAUM, GP .
RADIOLOGY, 1990, 175 (01) :91-96
[2]   TREATMENT OF EXPERIMENTALLY INDUCED ATHEROSCLEROSIS IN SWINE ILIAC ARTERIES - A COMPARISON OF SELF-EXPANDING AND BALLOON-EXPANDED STENTS [J].
CHUAPETCHARASOPON, C ;
WRIGHT, KC ;
WALLACE, S ;
DOBBEN, RL ;
GIANTURCO, C .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1992, 15 (03) :143-150
[3]   VENA TECH-LGM FILTER - LONG-TERM RESULTS OF A PROSPECTIVE-STUDY [J].
CROCHET, DP ;
STORA, O ;
FERRY, D ;
GROSSETETE, R ;
LEURENT, B ;
BRUNEL, P ;
NGUYEN, JM .
RADIOLOGY, 1993, 188 (03) :857-860
[4]   PERCUTANEOUS INFERIOR VENA-CAVAL FILTERS - FOLLOW-UP OF 7 DESIGNS IN 320 PATIENTS [J].
FERRIS, EJ ;
MCCOWAN, TC ;
CARVER, DK ;
MCFARLAND, DR .
RADIOLOGY, 1993, 188 (03) :851-856
[5]   STAGING OF FIXATION AND RETRIEVABILITY OF GREENFIELD FILTERS [J].
GREENFIELD, LJ ;
PROCTOR, MC ;
JAMES, EA ;
ABRAMS, GD ;
MOURSI, MM .
JOURNAL OF VASCULAR SURGERY, 1994, 20 (05) :744-750
[6]   RETRIEVABLE IVC FILTER - PRELIMINARY IN-VITRO AND IN-VIVO EVALUATION [J].
IRIE, T ;
YAMAUCHI, T ;
MAKITA, K ;
KUSANO, S .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1995, 6 (03) :449-454
[7]   A NEW VENA-CAVAL FILTER FOR PERCUTANEOUS PLACEMENT AND RETRIEVAL - EXPERIMENTAL-STUDY [J].
LUND, G ;
RYSAVY, JA ;
SALOMONOWITZ, E ;
CRAGG, AH ;
KOTULA, F ;
CASTENEDAZUNIGA, WR ;
HUNTER, DW ;
COLEMAN, CC ;
AMPLATZ, K .
RADIOLOGY, 1984, 152 (02) :369-372
[8]  
MAASS D, 1985, J CARDIOVASC SURG, V26, P116
[9]  
Murphy T P, 1991, J Vasc Interv Radiol, V2, P107, DOI 10.1016/S1051-0443(91)72482-2
[10]  
Ricco J B, 1988, Ann Vasc Surg, V2, P242, DOI 10.1016/S0890-5096(07)60009-0