Aerosol electrostatics .1. Properties of fine powders before and after aerosolization by dry powder inhalers

被引:74
作者
Byron, PR [1 ]
Peart, J [1 ]
Staniforth, JN [1 ]
机构
[1] UNIV BATH, SCH PHARM & PHARMACOL, BATH BA2 7AY, AVON, ENGLAND
关键词
dry powder inhaler; aerosol; electrostatic charge; aerosolization; deposition;
D O I
10.1023/A:1012181818244
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Purpose. To evaluate the dependence of fine particle dose charge (EPD charge) generated from powder inhalers on physico-chemical properties of the inhalation powder, inhaler type, deaggregation mechanism, dose number and/or retained powder. Methods. Electrostatic charges were determined on micronized powders and aerosolized fine particle doses withdrawn from two, high efficiency, multidose powder inhalers, Turbohaler(TM) and prototype Dryhaler(TM). The behavior of terbutaline sulfate, budesonide, albuterol (sulfate and base), beclomethasone dipropionate and lactose was assessed before and after aerosolization. Results. Both inhalers conferred triboelectric FPD charges during aerosolization in the range -400 pC through +200 pC. Specific charges (charge/unit mass) on the fine particle doses of budesonide from Dryhaler were significantly less than those from Turbohaler (p < 0.01), Electrostatic charges on the potentially respirable cloud of terbutaline sulfate generated by Bricanyl Turbohaler were positive and/or negative and unpredictable. With Pulmicort Turbohaler, FPD charges on budesonide were always positive. Dryhaler was used to determine the chemical dependence of fine particle triboelectrification during the aerosolization of pure materials. A triboelectric series was constructed from the Dryhaler results ranking the powders from positive to negative as budesonide > lactose > albuterol sulfate > terbutaline sulfate greater than or equal to albuterol greater than or equal to beclomethasone dipropionate. Conclusions. While there was no evidence of FPD charge dependence upon dose number with either inhaler, FPD charges were dependent upon the powder under investigation, as well as the construction and deaggregation mechanism of the inhaler. The specific charge on the fine particle dose of budesonide from Turbohaler corresponded to approximately 200 electronic charges per particle, a value which is known to affect both total and regional aerosol deposition in the human lung. Electrostatic charge effects may be important determinants of aerosol behavior and should not be neglected.
引用
收藏
页码:698 / 705
页数:8
相关论文
共 22 条
[1]  
[Anonymous], 1966, HEALTH PHYS, V12, P173
[2]  
Balachandran W., 1991, Electrostatics 1991. Invited and Contributed Papers from the Eighth International Conference, P57
[3]   THE EFFECT OF DELAY, MULTIPLE ACTUATIONS AND SPACER STATIC CHARGE ON THE IN-VITRO DELIVERY OF BUDESONIDE FROM THE NEBUHALER [J].
BARRY, PW ;
OCALLAGHAN, C .
BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1995, 40 (01) :76-78
[4]  
BARRY PW, 1994, THORAX, V49, P402
[5]  
Berg E, 1996, RESP DRUG DELIVERY, P308
[6]   DRUG ABSORPTION FROM INHALATION AEROSOLS ADMINISTERED BY POSITIVE-PRESSURE VENTILATION .1. ADMINISTRATION OF A CHARACTERIZED, SOLID DISODIUM FLUORESCEIN AEROSOL UNDER A CONTROLLED RESPIRATORY REGIME TO THE BEAGLE DOG [J].
BYRON, PR ;
CLARK, AR .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1985, 74 (09) :934-938
[7]   The effect of handling techniques on electrostatic charge on spacer devices: A correlation with in vitro particle size analysis [J].
Dewsbury, NJ ;
Kenyon, CJ ;
Newman, SP .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1996, 137 (02) :261-264
[8]  
DRACO A, COMMUNICATION
[9]   MODELING THE EFFECT OF CHARGE ON SELECTIVE DEPOSITION OF PARTICLES IN A DISEASED LUNG USING AEROSOL BOLI [J].
HASHISH, AH ;
BAILEY, AG ;
WILLIAMS, TJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (12) :2247-2262
[10]  
HILL M, 1994, RESPIRATORY DRUG DEL, V4, P109