Precision of transfer of preoperative planning for oral implants based on cone-beam CT-scan images through a robotic drilling machine - An in vitro study

被引:99
作者
Fortin, T
Champleboux, G
Bianchi, S
Buatois, H
Coudert, FL
机构
[1] Dent Univ Lyon, Dept Oral Surg, Lyon, France
[2] Med Univ, Grenoble, France
[3] Univ Turin, Dept Radiol, Turin, Italy
关键词
accuracy; computed tomography; computer-aided surgery; cone beam; dental imaging; implant dentistry; surgical template;
D O I
10.1034/j.1600-0501.2002.130612.x
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
In this study, an image-guided system for oral implant placement was assessed. A specially designed mechanical tool has been elaborated to transfer the preoperative implant axis planned on 3-dimensional imagery into a surgical template by a numerically controlled drilling machine. The main drawback of image-guiding systems is the use of preoperative computed tomography, which is expensive and delivers high radiation doses. Therefore, in this study the image-guiding system was coupled with a cone-beam tomograph that significantly decreased both cost and radiation doses. Three edentulous models were used. To determine the accuracy of the system, the ability of a 1.8-mm diameter drill to enter a 2.0-mm diameter, 10-mm-long titanium tube inserted on the model with no contact was verified. Because the drill entered the tubes with no contact and went beyond the end of the tube, the transfer error was less than 0.2 mm for translation and less than 1.1degrees for rotation. The method presented here is low cost and high precision compared to other technological solutions such as tracking. Further assessment in the surgical field should lead to daily use of this system for flapless surgery, to prepare a prosthesis prior to surgery for immediate loading, to reduce risk of injuring critical anatomical structures and to eliminate manual placement error.
引用
收藏
页码:651 / 656
页数:6
相关论文
共 25 条
[1]
Development of a compact computed tomographic apparatus for dental use [J].
Arai, Y ;
Tammisalo, E ;
Iwai, K ;
Hashimoto, K ;
Shinoda, K .
DENTOMAXILLOFACIAL RADIOLOGY, 1999, 28 (04) :245-248
[2]
Prosthetic management of the partially dentate patient with fixed implant restorations [J].
Belser, UC ;
Mericske-Stern, R ;
Bernard, JP ;
Taylor, TD .
CLINICAL ORAL IMPLANTS RESEARCH, 2000, 11 :126-145
[3]
In-vitro assessment of a registration protocol for image guided implant dentistry [J].
Birkfellner, W ;
Solar, P ;
Gahleitner, A ;
Huber, K ;
Kainberger, F ;
Kettenbach, J ;
Homolka, P ;
Diemling, M ;
Watzek, G ;
Bergmann, H .
CLINICAL ORAL IMPLANTS RESEARCH, 2001, 12 (01) :69-78
[4]
Brief J, 2000, INT CONGR SER, V1214, P932
[5]
COMPUTER-ASSISTED MEDICAL INTERVENTIONS - PASSIVE AND SEMIACTIVE AIDS [J].
CINQUIN, P ;
BAINVILLE, E ;
BARBE, C ;
BITTAR, E ;
BOUCHARD, V ;
BRICAULT, I ;
CHAMPLEBOUX, G ;
CHENIN, M ;
CHEVALIER, L ;
DELNONDEDIEU, Y ;
DESBAT, L ;
DESSENNE, V ;
HAMADEH, A ;
HENRY, D ;
LAIEB, N ;
LAVALLEE, S ;
LEFEBVRE, JM ;
LEITNER, F ;
MENGUY, Y ;
PADIEU, F ;
PERIA, O ;
POYET, A ;
PROMAYON, M ;
ROUAULT, S ;
SAUTOT, P ;
TROCCAZ, J ;
VASSAL, P .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1995, 14 (03) :254-263
[6]
Demey S, 1999, INT CONGR SER, V1191, P883
[7]
OPTIMUM PLACEMENT OF OSSEOINTEGRATED IMPLANTS [J].
ENGELMAN, MJ ;
SORENSEN, JA ;
MOY, P .
JOURNAL OF PROSTHETIC DENTISTRY, 1988, 59 (04) :467-473
[8]
Fortin T, 2000, J Oral Implantol, V26, P300, DOI 10.1563/1548-1336(2000)026<0300:PDIPIB>2.3.CO
[9]
2
[10]
Fortin T, 1995, J Image Guid Surg, V1, P53, DOI 10.1002/(SICI)1522-712X(1995)1:1<53::AID-IGS8>3.3.CO