An on-line system for preconcentration and determination of lead in wine samples by FAAS

被引:87
作者
Lemos, VA
de la Guardia, M
Ferreira, SLC
机构
[1] Univ Estadual Sudoeste Bahia, Lab Quim Analit, BR-45200000 Jequie, BA, Brazil
[2] Univ Valencia, Dept Quim Analyt, Valencia, Spain
[3] Univ Fed Bahia, Grp Pesquisa Quim Analit, Inst Quim, BR-40170290 Salvador, BA, Brazil
关键词
lead; wine; FAAS; on-line;
D O I
10.1016/S0039-9140(02)00325-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection method has been developed for the direct determination of free available Pb(II) and total Pb content in wine samples. The method is based on the chemical sorption of Pb(11), from pH 7 buffered solutions, on a packed polyurethane foam column, modified by addition of 2-(2-benzothiazolylazo)-p-cresol (BTAC). After this step, lead was directly eluted with a stream of 0.1 mol l(-1) HCl into an air C2H2 flame in which lead was determined by atomic absorption spectrometry. Total lead was analyzed after sample digestion with nitric acid and hydrogen peroxide, being free available lead determined by direct sample on-line preconcentration and elution. The method provides a limit of detection (3sigma) of I mug l(-1) lead and relative standard deviation, which varies from 6 to 0.7% for lead concentration of 10 and 500 mug l(-1). Total content of lead in wine samples analyzed varied from 8 to 42 mug l(-1) being obtained free available values of Pb(II) under the limit of detection of the method. Recovery studies on natural wine samples, spicked with inorganic lead, evidenced the remaining capability of ligands, present in the wine, to avoid lead retention on the polyurethane foam loaded with BTAC. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:475 / 480
页数:6
相关论文
共 26 条
[1]  
Almeida CMR, 1999, ANAL CHIM ACTA, V396, P45
[2]   UV-irradiation and MW-digestion pre-treatment of Port wine suitable for the determination of lead isotope ratios by inductively coupled plasma mass spectrometry [J].
Almeida, CMR ;
Vasconcelos, MTSD .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (12) :1815-1821
[3]   Pb and Cu speciation and bioavailability in port wine [J].
Azenha, MAGO ;
Vasconcelos, MTSD .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2000, 48 (11) :5740-5749
[4]  
BALDO MA, 1997, ANALYST, V1, P122
[5]   Evaluation of the accuracy of the determination of lead isotope ratios in wine by ICP MS using quadrupole, multicollector magnetic sector and time-of-flight analyzers [J].
Barbaste, M ;
Halicz, L ;
Galy, A ;
Medina, B ;
Emteborg, H ;
Adams, FC ;
Lobinski, R .
TALANTA, 2001, 54 (02) :307-317
[6]   ABSORPTION BY POLYURETHANE FOAMS - NEW METHOD OF SEPARATION [J].
BOWEN, HJM .
JOURNAL OF THE CHEMICAL SOCIETY A -INORGANIC PHYSICAL THEORETICAL, 1970, (07) :1082-&
[7]   DETERMINATION OF LEAD IN WINE, OTHER BEVERAGES AND FRUIT SLURRIES BY FLOW-INJECTION HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY WITH ONLINE MICROWAVE DIGESTION [J].
CABRERA, C ;
MADRID, Y ;
CAMARA, C .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (12) :1423-1426
[8]   Selectivity enhancement in spectrophotometry: on-line interference suppression using polyurethane foam minicolumn for aluminum determination with Methyl Thymol Blue [J].
Cassella, RJ ;
Santelli, RE ;
Branco, AG ;
Lemos, VA ;
Ferreira, SLC ;
de Carvalho, MS .
ANALYST, 1999, 124 (05) :805-808
[9]  
Clarkson TW, 1988, BIOL MONITORING TOXI
[10]  
Fang Z., 1993, FLOW INJECTION SEPAR