Cement mantle stress under retroversion torque at heel-strike

被引:9
作者
Afsharpoya, B. [1 ]
Barton, D. C. [1 ]
Fisher, J. [1 ]
Purbach, B. [2 ]
Wroblewski, M. [2 ]
Stewart, T. D. [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Wrightington Hosp, Wigan, England
关键词
Stress; Hip prostheses; Cement; Fixation; TOTAL HIP-ARTHROPLASTY; FEMORAL COMPONENTS; PROXIMAL FEMUR; FAILURE; MODEL; STEM; REPLACEMENTS; PROSTHESES; INTERFACE; MUSCLE;
D O I
10.1016/j.medengphy.2009.09.006
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
The paper presents a theory of fixation failure and loosening in cemented total hip prostheses and proceeds to investigate this using an experimentally validated finite element model and two prosthesis types, namely the Charnley and the C-Stem. The study investigates the effects of retroversion torque occurring at heel-strike in combination with a loss of proximal cement/bone support and distal implant/cement support with a good distal cement/bone interface. A 3D finite element model was validated by comparison of femoral surface strains with those measured in an in vitro experimental simulation using an implanted Sawbone femur loaded in the heel-strike position and including a simplified representation of muscle forces. Results showed that the heel-strike position applies a high retroversion torque to the femoral stem that when combined with proximal debonding of the cement/bone interface and distal debonding of the implant/cement interface increases the strain transfer to the cement that may ultimately lead to the breakdown of the cement mantle leading on to osteolysis and loosening of the prostheses. Experimental fatigue testing of the implanted Charnely stem in a Sawbone femur produced cracks within the cement mantle that were located in positions of maximum stress supporting the finite element analysis results and theory of failure. (C) 2009 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1323 / 1330
页数:8
相关论文
共 33 条
[1]
AFSHARPOYA B, 2002, EXPT FINITE ELEMENT
[2]
[Anonymous], 1979, Clin Orthop Relat Res
[3]
[Anonymous], 7206 ISO
[4]
[Anonymous], CLIN ORTHOPAEDICS
[5]
HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[6]
Reproduction of fretting wear at the stem-cement interface in total hip replacement [J].
Brown, L. ;
Zhang, H. ;
Blunt, L. ;
Barrans, S. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2007, 221 (H8) :963-971
[7]
ROLE OF LOADS AND PROSTHESIS MATERIAL PROPERTIES ON THE MECHANICS OF THE PROXIMAL FEMUR AFTER TOTAL HIP-ARTHROPLASTY [J].
CHEAL, EJ ;
SPECTOR, M ;
HAYES, WC .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1992, 10 (03) :405-422
[8]
INFLUENCE OF THIGH MUSCLES ON THE AXIAL STRAINS IN A PROXIMAL FEMUR DURING EARLY STANCE IN GAIT [J].
CRISTOFOLINI, L ;
VICECONTI, M ;
TONI, A ;
GIUNTI, A .
JOURNAL OF BIOMECHANICS, 1995, 28 (05) :617-624
[9]
DARMON L, 2001, 47 ANN M ORTH RES SO, P1068
[10]
A 3-DIMENSIONAL BIOMECHANICAL MODEL OF HIP MUSCULATURE [J].
DOSTAL, WF ;
ANDREWS, JG .