Estimation of mean-glandular dose from monitoring breast entrance skin air kerma using a high sensitivity metal oxide semiconductor field effect transistor (MOSFET) dosimeter system in mammography

被引:19
作者
Dong, SL
Chu, TC
Lee, JS
Lan, GY
Wu, TH
Yeh, YH
Hwang, JJ
机构
[1] Natl Yang Ming Univ, Inst Radiol Sci, Taipei, Taiwan
[2] Natl Tsing Hua Univ, Dept Nucl Sci, Hsinchu, Taiwan
[3] Yuan Pei Inst Sci & Technol, Dept Radiol Technol, Taipei 112, Taiwan
关键词
MOSFET dosimeter; mean-glandular dose; BESAK;
D O I
10.1016/S0969-8043(02)00172-0
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Estimation of mean-glandular dose (MGD) has been investigated in recent years due to the potential risks of radiation-induced carcinogenesis associated with the mammographic examination for diagnostic radiology. In this study, a new technique for immediate readout of breast entrance skin air kerma (BESAK) using high sensitivity MOSFET dosimeter after mammographic projection was introduced and a formula for the prediction of tube output with exposure records was developed. A series of appropriate conversion factors was applied to the MGD determination from the BESAK. The study results showed that signal response of the high sensitivity MOSFET exhibited excellent linearity within mammographic dose ranges, and that the energy dependence was less than 3% for each anode/filter combination at the tube potentials 25-30 kV. Good agreement was observed between the BESAK and the tube exposure output measurement for breasts thicker than 30 mm. In addition, the air kerma estimated from our prediction formula provided sufficient accuracy for thinner breasts. The average MGD from 120 Asian females was 1.5 mGy, comparable to other studies. Our results suggest that the high sensitivity MOSFET dosimeter system is a good candidate for immediately readout of BESAK after mammographic procedures. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:791 / 799
页数:9
相关论文
共 21 条
[1]   Dosimetric implications of age related glandular changes in screening mammography [J].
Beckett, JR ;
Kotre, CJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (03) :801-813
[2]   The characterization of a commercial MOSFET dosimeter system for use in diagnostic x ray [J].
Bower, MW ;
Hintenlang, DE .
HEALTH PHYSICS, 1998, 75 (02) :197-204
[3]  
BULLING SM, 1995, RADIAT PROT DOSIM, V57, P455
[4]   A new radiotherapy surface dose detector: The MOSFET [J].
Butson, MJ ;
Rozenfeld, A ;
Mathur, JN ;
Carolan, M ;
Wong, TPY ;
Metcalfe, PE .
MEDICAL PHYSICS, 1996, 23 (05) :655-658
[5]   Breast dosimetry [J].
Dance, DR ;
Skinner, CL ;
Carlsson, GA .
APPLIED RADIATION AND ISOTOPES, 1999, 50 (01) :185-203
[6]   Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol [J].
Dance, DR ;
Skinner, CL ;
Young, KC ;
Beckett, JR ;
Kotre, CJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3225-3240
[7]   MONTE-CARLO CALCULATION OF CONVERSION FACTORS FOR THE ESTIMATION OF MEAN GLANDULAR BREAST DOSE [J].
DANCE, DR .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1211-1219
[8]  
DONG SL, IN PRESS APPL RAD IS
[9]   ASSESSMENT OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY [J].
FAULKNER, K ;
LAW, J ;
ROBSON, KJ .
BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (812) :877-881
[10]   TLD measurements of in vivo mammographic exposures and the calculated mean glandular dose across the United States [J].
Gentry, JR ;
DeWerd, LA .
MEDICAL PHYSICS, 1996, 23 (06) :899-903