Estimation of mean-glandular dose from monitoring breast entrance skin air kerma using a high sensitivity metal oxide semiconductor field effect transistor (MOSFET) dosimeter system in mammography

被引:19
作者
Dong, SL
Chu, TC
Lee, JS
Lan, GY
Wu, TH
Yeh, YH
Hwang, JJ
机构
[1] Natl Yang Ming Univ, Inst Radiol Sci, Taipei, Taiwan
[2] Natl Tsing Hua Univ, Dept Nucl Sci, Hsinchu, Taiwan
[3] Yuan Pei Inst Sci & Technol, Dept Radiol Technol, Taipei 112, Taiwan
关键词
MOSFET dosimeter; mean-glandular dose; BESAK;
D O I
10.1016/S0969-8043(02)00172-0
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Estimation of mean-glandular dose (MGD) has been investigated in recent years due to the potential risks of radiation-induced carcinogenesis associated with the mammographic examination for diagnostic radiology. In this study, a new technique for immediate readout of breast entrance skin air kerma (BESAK) using high sensitivity MOSFET dosimeter after mammographic projection was introduced and a formula for the prediction of tube output with exposure records was developed. A series of appropriate conversion factors was applied to the MGD determination from the BESAK. The study results showed that signal response of the high sensitivity MOSFET exhibited excellent linearity within mammographic dose ranges, and that the energy dependence was less than 3% for each anode/filter combination at the tube potentials 25-30 kV. Good agreement was observed between the BESAK and the tube exposure output measurement for breasts thicker than 30 mm. In addition, the air kerma estimated from our prediction formula provided sufficient accuracy for thinner breasts. The average MGD from 120 Asian females was 1.5 mGy, comparable to other studies. Our results suggest that the high sensitivity MOSFET dosimeter system is a good candidate for immediately readout of BESAK after mammographic procedures. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:791 / 799
页数:9
相关论文
共 21 条
[11]  
*ICRP, 1991, ANN ICRP, V21, P5
[12]  
*IPSM, 1994, 59 IPSM
[13]   Determination of average glandular dose with modern mammography units for two large groups of patients [J].
Klein, R ;
Aichinger, H ;
Dierker, J ;
Jansen, JTM ;
JoiteBarfuss, S ;
Sabel, M ;
SchulzWendtland, R ;
Zoetelief, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (04) :651-671
[14]   COMPARATIVE-STUDY OF DOSE VALUES AND IMAGE QUALITY IN MAMMOGRAPHY IN THE AREA OF MADRID [J].
MORAN, P ;
CHEVALIER, M ;
VANO, E .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (798) :556-563
[15]   Measurement of radiotherapy x-ray skin dose on a chest wall phantom [J].
Quach, KY ;
Morales, J ;
Butson, MJ ;
Rosenfeld, AB ;
Metcalfe, PE .
MEDICAL PHYSICS, 2000, 27 (07) :1676-1680
[16]   Clinical dosimetry using MOSFETS [J].
Ramani, R ;
Russell, S ;
OBrien, P .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (04) :959-964
[17]  
Rosenstein M., 1985, HHS PUBLICATION FDA, P85
[18]  
THILANDER A, 1992, RADIAT PROT DOSIM, V43, P217
[19]  
WALL MA, 1992, RADIAT PROT DOSIM, V43, P253
[20]   SPECTRAL DEPENDENCE OF GLANDULAR TISSUE DOSE IN SCREEN-FILM MAMMOGRAPHY [J].
WU, XZ ;
BARNES, GT ;
TUCKER, DM .
RADIOLOGY, 1991, 179 (01) :143-148