Radiation dose optimization in coronary angiography and percutaneous coronary intervention (PCI). I. Experimental studies

被引:12
作者
Geijer, H [1 ]
Beckman, KW
Andersson, T
Persliden, J
机构
[1] Orebro Univ Hosp, Dept Radiol, S-70185 Orebro, Sweden
[2] Orebro Univ Hosp, Dept Med Phys, S-70185 Orebro, Sweden
关键词
coronary intervention; coronary angiography; coronary vessels; transluminal angioplasty; radiation exposure to patients; diagnostic radiology;
D O I
10.1007/s00330-001-1237-6
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The objectives of this study were to evaluate the influence on image quality and dose to the patient and operator of various equipment settings for percutaneous coronary intervention (PCI), and to optimize the set-up. With an Alderson phantom, different settings, such as projection, protective screens, filtration, image intensifier size and collimation, were evaluated. Kerma-area product (KAP) was recorded as a measure of patient dose and scattered radiation was measured with an ionization chamber. Effective dose for a standardized PCI procedure was measured with thermoluminescent dosimeters inside the phantom. Image quality was evaluated with a contrast-detail phantom. Based on these findings, the equipment set-up was optimized to a low fluoroscopy dose rate with a sufficient image quality. Several operating parameters affected dose, particularly scattered radiation. The optimization reduced the fluoroscopy KAP rate from 44 to 16 mGy cm(2)/s using 15 cm of acrylic. The effective dose was reduced from 13 to 4.6 mSv for a standardized PCI procedure. Radiation dose to patient and operator in PCI is heavily dependent on both equipment set-up and operating parameters which can be influenced by the operator. With a careful optimization, a large reduction of radiation dose is possible.
引用
收藏
页码:2571 / 2581
页数:11
相关论文
共 40 条
[1]  
ALDERSON SW, 1962, AMER J ROENTGENOL RA, V87, P185
[2]  
[Anonymous], 1990, ICRP PUBL
[3]  
[Anonymous], ICRP PUBL
[4]   Patient radiation doses during cardiac catheterization procedures [J].
Betsou, S ;
Efstathopoulos, EP ;
Katritsis, D ;
Faulkner, K ;
Panayiotakis, G .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (846) :634-639
[5]   Dose reduction in gastrointestinal and genitourinary fluoroscopy: Use of grid-controlled pulsed fluoroscopy [J].
Boland, GWL ;
Murphy, B ;
Arellano, R ;
Niklason, L ;
Mueller, PR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (05) :1453-1457
[6]   The impact of cardiology on the collective effective dose in the North of England [J].
Broadhead, DA ;
Chapple, CL ;
Faulkner, K ;
Davies, ML ;
McCallum, H .
BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (833) :492-497
[7]   THE CASE FOR RADIOPROTECTIVE EYEWEAR FACEWEAR - PRACTICAL IMPLICATIONS AND SUGGESTIONS [J].
COUSIN, AJ ;
LAWDAHL, RB ;
CHAKRABORTY, DP ;
KOEHLER, RE .
INVESTIGATIVE RADIOLOGY, 1987, 22 (08) :688-692
[8]   Real-time measurement of radiation exposure to patients during diagnostic coronary angiography and percutaneous interventional procedures [J].
Cusma, JT ;
Bell, MR ;
Wondrow, MA ;
Taubel, JP ;
Holmes, DR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 33 (02) :427-435
[9]   REDUCTION OF RADIATION EXPOSURE WHILE MAINTAINING HIGH-QUALITY FLUOROSCOPIC IMAGES DURING INTERVENTIONAL CARDIOLOGY USING NOVEL X-RAY TUBE TECHNOLOGY WITH EXTRA BEAM FILTERING [J].
DENBOER, A ;
DEFEYTER, PJ ;
HUMMEL, WA ;
KEANE, D ;
ROELANDT, JRTC .
CIRCULATION, 1994, 89 (06) :2710-2714
[10]   Patient radiation exposure during coronary angiography and intervention [J].
Fransson, SG ;
Persliden, J .
ACTA RADIOLOGICA, 2000, 41 (02) :142-144