Electrochemical immobilisation of enzymes on conducting organic salt electrodes: characterisation of an oxygen independent and interference-free glucose biosensor

被引:30
作者
Centonze, D
Losito, I
Malitesta, C
Palmisano, F
Zambonin, PG
机构
[1] UNIV BARI,DIPARTMENTO CHIM,I-70126 BARI,ITALY
[2] UNIV BASILICATA,DIPARTIMENTO CHIM,I-85100 POTENZA,ITALY
来源
JOURNAL OF ELECTROANALYTICAL CHEMISTRY | 1997年 / 435卷 / 1-2期
关键词
glucose biosensor; electrochemical immobilisation; conducting organic salt;
D O I
10.1016/S0022-0728(97)00294-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A glucose biosensor, based on the electrochemical immobilisation of glucose oxidase (GOD) in a poly(o-phenylendiamine) (PPD) film synthesised onto a NMP TCNQ conducting organic salt (COS) electrode, is described and its performances are evaluated. The electrochemical cycling of the bare conducting organic salt electrode, in a phosphate buffer (pH 7) supporting electrolyte, accumulated a surface layer of TCNQ(0), which was supposed to be the active mediator for the heterogeneous reoxidation of GOD. Under certain boundary conditions, Chen et al.'s model [C.J. Chen, C.C. Liu, R. Savinell, J. Electroanal. Chem., 248 (1993) 317] has been applied to describe the catalysis occurring at our biosensor. Moreover, when the electrolysis is fast, such a model was demonstrated to be equivalent to Albery et al.'s mathematics [W.J. Albery, P.N. Bartlett, D.H. Craston. J. Electroanal Chem., 194 (1985) 223], which allows the estimation of the kinetic constants. Such parameters indicated:hat substrate diffusion through the membrane and unsaturated enzyme kinetics were the rate-limiting processes. COS/PPD/GOD electrode respond rapidly to glucose, the steady-state current being reached in 10-16 s. Biosensor responses under different operating conditions such as applied potential, presence of oxygen or interferents were investigated. The sensor response, tested in a day to-day experiment, remained quite stable for 4 days, then decreased to 60% of the initial value on day 5. The shelf lifetime was at least 1 month. (C) 1997 Elsevier Science S.A.
引用
收藏
页码:103 / 111
页数:9
相关论文
共 49 条
[21]   ENZYME ELECTROCATALYSIS AT ORGANIC SALT ELECTRODES [J].
HILL, BS ;
SCOLARI, CA ;
WILSON, GS .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1988, 252 (01) :125-138
[22]   IMMOBILIZED XANTHINE-OXIDASE CHEMICALLY MODIFIED ELECTRODE AS A DUAL ANALYTICAL SENSOR [J].
IANNIELLO, RM ;
LINDSAY, TJ ;
YACYNYCH, AM .
ANALYTICAL CHEMISTRY, 1982, 54 (12) :1980-1984
[23]   GLUCOSE-OXIDASE IMMOBILIZED BENZOQUINONE CARBON PASTE ELECTRODE AS A GLUCOSE SENSOR [J].
IKEDA, T ;
HAMADA, H ;
MIKI, K ;
SENDA, M .
AGRICULTURAL AND BIOLOGICAL CHEMISTRY, 1985, 49 (02) :541-543
[24]   ELECTROCHEMICAL-BEHAVIOR OF DONOR-TETRACYANOQUINODIMETHANE ELECTRODES IN AQUEOUS-MEDIA [J].
JAEGER, CD ;
BARD, AJ .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1980, 102 (17) :5435-5442
[25]   ELECTROCHEMICAL BEHAVIOR OF TETRATHIAFULVALENE-TETRACYANOQUINODIMETHANE ELECTRODES IN AQUEOUS-MEDIA [J].
JAEGER, CD ;
BARD, AJ .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1979, 101 (07) :1690-1699
[26]   IMPEDANCE ANALYSIS OF NMP-TCNQ AT GLUCOSE-OXIDASE INTERFACES [J].
JUNG, CC ;
HALL, EAH .
ANALYTICAL CHEMISTRY, 1995, 67 (14) :2393-2400
[27]  
KADISH AH, 1968, CLIN CHEM, V14, P116
[28]   GLUCOSE SENSITIVITY OF POLYPYRROLE FILMS CONTAINING IMMOBILIZED GLUCOSE-OXIDASE AND HYDROQUINONESULFONATE IONS [J].
KAJIYA, Y ;
SUGAI, H ;
IWAKURA, C ;
YONEYAMA, H .
ANALYTICAL CHEMISTRY, 1991, 63 (01) :49-54
[29]   ENZYME-MODIFIED ORGANIC CONDUCTING SALT MICROELECTRODE [J].
KAWAGOE, JL ;
NIEHAUS, DE ;
WIGHTMAN, RM .
ANALYTICAL CHEMISTRY, 1991, 63 (24) :2961-2965
[30]   KINETIC-STUDY OF THE PERFORMANCE OF 3RD-GENERATION BIOSENSORS [J].
KOOPAL, CGJ ;
NOLTE, RJM .
BIOELECTROCHEMISTRY AND BIOENERGETICS, 1994, 33 (01) :45-53