Carotid Artery Stents: In Vitro Comparison of Different Stent Designs and Sizes Using CT Angiography and Contrast-Enhanced MR Angiography at 15T and 3T

被引:22
作者
Lettau, M. [1 ]
Sauer, A. [1 ]
Heiland, S. [1 ]
Rohde, S. [1 ]
Bendszus, M. [1 ]
Haehnel, S. [1 ]
机构
[1] Univ Heidelberg, Med Ctr, Dept Neurol, Div Neuroradiol, D-69120 Heidelberg, Germany
关键词
MAGNETIC-RESONANCE ANGIOGRAPHY; METALLIC VASCULAR IMPLANTS; LUMEN VISUALIZATION; ILIAC ARTERIES; 3.0; TESLA; ARTIFACTS; SUSCEPTIBILITY; COMPATIBILITY; ANGIOPLASTY; APPEARANCE;
D O I
10.3174/ajnr.A1743
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND AND PURPOSE: CT angiography (CTA) and MR angiography (MRA) are increasingly used methods for evaluation of stented vessel segments. Our aim was to compare CTA, contrast-enhanced MRA (CE-MRA) at 1.5T, and CE-MRA at 3T for the visualization of carotid artery stents and to define the best noninvasive imaging technique as an alternative to conventional angiography for each stent. MATERIALS AND METHODS: CTA and CE-MRA appearances of 18 carotid artery stents of different designs and sizes (4.0 to 10.0 mm) were investigated in vitro. For each stent, artificial lumen narrowing (ALN) was calculated. RESULTS: With CE-MRA at 3T and at 1.5T, ALN in most nitinol stents was lower than that in the groups of stainless steel and cobalt alloy stents. In most nitinol stents and in both cobalt alloy stents, ALN was lower on CE-MRA at 3T than at 1.5T. In all stainless steel stents, ALN was lower on CTA than on CE-MRA. With CTA and CE-MRA, in most stents ALN decreased with increasing stent diameter. CONCLUSIONS: CTA and CE-MRA evaluation of vessel patency after stent placement is possible but is considerably impaired by ALN. Investigators should be informed about the method of choice for every stent. Stent manufacturers should be aware of potential artifacts caused by their stents during noninvasive diagnostic methods such as CTA and CE-MRA.
引用
收藏
页码:1993 / 1997
页数:5
相关论文
共 26 条
[1]   Metallic artifacts of coronary and iliac arteries stents in mr angiography and contrast-enhanced CT [J].
Amano, Y ;
Ishihara, M ;
Hayashi, H ;
Gemma, K ;
Kawamata, H ;
Amano, M ;
Kumazaki, T .
CLINICAL IMAGING, 1999, 23 (02) :85-89
[2]   SIMULATION OF SUSCEPTIBILITY ARTIFACTS IN 2D AND 3D FOURIER-TRANSFORM SPIN-ECHO AND GRADIENT-ECHO MAGNETIC-RESONANCE-IMAGING [J].
BAKKER, CJG ;
BHAGWANDIEN, R ;
MOERLAND, MA ;
RAMOS, LMP .
MAGNETIC RESONANCE IMAGING, 1994, 12 (05) :767-774
[3]   Improved lumen visualization in metallic vascular implants by reducing RF artifacts [J].
Bartels, LW ;
Bakker, CJG ;
Viergever, MA .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (01) :171-180
[4]   MR imaging of vascular stents: effects of susceptibility, flow, and radiofrequency eddy currents [J].
Bartels, LW ;
Smits, HFM ;
Bakker, CJG ;
Viergever, MA .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2001, 12 (03) :365-371
[5]   Silent embolism in diagnostic cerebral angiography and neurointerventional procedures: a prospective study [J].
Bendszus, M ;
Koltzenburg, M ;
Burger, R ;
Warmuth-Metz, M ;
Hofmann, E ;
Solymosi, L .
LANCET, 1999, 354 (9190) :1594-1597
[6]  
Bladin CF, 1996, INT ANGIOL, V15, P295
[7]   Artifact-free in-stent lumen visualization by standard magnetic resonance angiography using a new metallic magnetic resonance imaging stent [J].
Buecker, A ;
Spuentrup, E ;
Ruebben, A ;
Günther, RW .
CIRCULATION, 2002, 105 (15) :1772-1775
[8]   Results of the Stent-Protected Angioplasty versus Carotid Endarterectomy (SPACE) study to treat symptomatic stenoses at 2 years: a multinational, prospective, randomised trial (vol 7, pg 893, 2008) [J].
Eckstein, H-H ;
Ringleb, P. ;
Allenberg, J-R .
LANCET NEUROLOGY, 2009, 8 (02) :135-135
[9]  
Graf H, 2003, ROFO-FORTSCHR RONTG, V175, P1711
[10]   3.0 Tesla contrast-enhanced mr angiography of carotid artery stents:: In vitro measurements as compared with 1.5 Tesla [J].
Haehnel, S. ;
Nguyen-Trong, T. H. ;
Rohde, S. ;
Hartmann, M. ;
Braun, C. ;
Sartor, K. ;
Heiland, S. .
JOURNAL OF NEURORADIOLOGY, 2006, 33 (02) :75-80