A Monte Carlo investigation of fluence profiles collimated by an electron specific MLC during beam delivery for modulated electron radiation therapy

被引:15
作者
Deng, J
Lee, MC
Ma, CM
机构
[1] Yale Univ, Sch Med, Dept Therapeut Radiol, New Haven, CT 06510 USA
[2] Stanford Univ, Sch Med, Dept Radiat Oncol, Stanford, CA 94305 USA
[3] Fox Chase Canc Ctr, Dept Radiat Oncol, Philadelphia, PA 19111 USA
关键词
electron beams; multileaf collimator; beam collimation; EGS4 Monte Carlo;
D O I
10.1118/1.1513160
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Modulated electron radiation therapy (MERT) is able to deliver conformal dose to shallow tumors while significantly reducing dose to distal structures and surrounding tissues. An electron specific multileaf collimator (eMLC) has been proposed and constructed as an effective means of delivering electron beams for MERT. The aim of this work is to apply the Monte Carlo method to investigate the fluence profiles collimated by the eMLC in order to achieve accurate beam delivery for MERT. In this work, the EGS4/BEAM code was used to simulate the eMLC collimated electron beams of 6-20 MeV generated from a Varian Clinac 2100C linear accelerator. An attempt was made to describe the fluence profiles with an analytic Sigmoid function. The function parameters were determined by the fittings of the Monte Carlo simulated fluence profiles. How the function parameters depend on the eMLC aperture size, the off-axis location, and the electron beam energy has been investigated. It has been found that the eMLC collimated fluence profiles are dependent on beam energy, while almost independent of leaf location or dimension of MLC aperture. There is little difference in the fluence profiles collimated by the leaf side and the leaf end for the straight-edged leaves. It is possible that these energy-dependent Sigmoid functions can serve as operators to account for the energy dependence of the eMLC collimated fluence profiles. These operators can be incorporated into the inverse planning algorithm to derive desired dose distributions using a set of electron beams of variable energy and field size suitable for delivery by the eMLC. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:2472 / 2483
页数:12
相关论文
共 23 条
[1]  
Asell M, 1997, PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, P47
[2]   Optimization of 3D conformal electron beam therapy in inhomogeneous media by concomitant fluence and energy modulation [J].
Asell, M ;
Hyodynmaa, S ;
Gustafsson, A ;
Brahme, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11) :2083-2100
[3]  
BIELAJEW AF, 1987, NUCL INSTRUM METH B, V18, P165
[4]   DESIGN PRINCIPLES AND CLINICAL POSSIBILITIES WITH A NEW GENERATION OF RADIATION-THERAPY EQUIPMENT - A REVIEW [J].
BRAHME, A .
ACTA ONCOLOGICA, 1987, 26 (06) :403-412
[5]   Possibilities for tailoring dose distributions through the manipulation of electron beam characteristics [J].
Ebert, MA ;
Hoban, PW .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11) :2065-2081
[6]   A MONTE-CARLO INVESTIGATION OF ELECTRON-BEAM APPLICATOR SCATTER [J].
EBERT, MA ;
HOBAN, PW .
MEDICAL PHYSICS, 1995, 22 (09) :1431-1435
[7]   A MODEL FOR ELECTRON-BEAM APPLICATOR SCATTER [J].
EBERT, MA ;
HOBAN, PW .
MEDICAL PHYSICS, 1995, 22 (09) :1419-1429
[8]   Optimization of conformal electron beam therapy using energy- and fluence-modulated beams [J].
Hyodynmaa, S ;
Gustafsson, A ;
Brahme, A .
MEDICAL PHYSICS, 1996, 23 (05) :659-666
[9]  
Jiang S. B., 2000, International Journal of Radiation Oncology Biology Physics, V48, P218, DOI 10.1016/S0360-3016(00)80230-8
[10]   Monte Carlo calculations of electron beam output factors for a medical linear accelerator [J].
Kapur, A ;
Ma, CM ;
Mok, EC ;
Findley, DO ;
Boyer, AL .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (12) :3479-3494