Real-time studies of the pivot bearings in the NEDO gyro PI-710 centrifugal blood pump

被引:9
作者
Asai, T
Watanabe, K
Ito, S
Tsujimura, S
Motomura, T
Shinohara, T
Glueck, JA
Nosé, Y
机构
[1] Baylor Coll Med, Michael E DeBakey Dept Surg, Houston, TX 77030 USA
[2] Miwatec Co Ltd, Kanagawa, Japan
关键词
centrifugal pump; long-term ventricular assist device; pivot bearing; ceramic; ultrahigh molecular weight polyethylene;
D O I
10.1111/j.1525-1594.2004.00074.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The NEDO Gyro PI-710 centrifugal pump (Gyro PI-710 pump) incorporates a double pivot bearing system of which the male pivot and female bearings are fabricated from Al2O3 ceramic and ultrahigh molecular weight polyethylene. The top female bearing is a critical component because the impeller is levitated by hydraulic force and is maintained in the top contact position. A long-term in vitro examination of the pivot bearings was conducted using a biventricular assist model. In 7 animal experiments, the depth change of the top female bearing was examined. Animal experiments up to 90 days revealed that there was no noticeable depth increase in the top female bearing. According to the in vitro study, the life of the pivot bearings of the left and right pump was estimated to be approximately 3 and 7 years, respectively. Further improvement of this pivot bearing system is currently underway.
引用
收藏
页码:899 / 903
页数:5
相关论文
共 9 条
[1]
HeartMate III: Pump design for a centrifugal LVAD with a magnetically levitated rotor [J].
Bourque, K ;
Gernes, DB ;
Loree, HM ;
Richardson, JS ;
Poirier, VL ;
Barletta, N ;
Fleischli, A ;
Foiera, G ;
Gempp, TM ;
Schoeb, R ;
Litwak, KN ;
Akimoto, T ;
Watach, MJ ;
Litwak, P .
ASAIO JOURNAL, 2001, 47 (04) :401-405
[2]
In vivo evaluation of the NEDO biventricular assist device with an RPM dynamic impeller suspension system [J].
Ichikawa, S ;
Linneweber, J ;
Motomura, T ;
Ishitoya, H ;
Watanabe, K ;
Ashizawa, S ;
Murai, N ;
Nishimura, I ;
Sumikura, H ;
Glueck, JA ;
Shinohara, T ;
Oestmann, DJ ;
Nosé, Y .
ASAIO JOURNAL, 2003, 49 (05) :578-582
[3]
Evaluation of the wear of the pivot bearing in the Gyro C1E3 pump [J].
Makinouchi, K ;
Nakazawa, T ;
Takami, Y ;
Takatani, S ;
Nose, Y .
ARTIFICIAL ORGANS, 1996, 20 (06) :523-528
[4]
Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls [J].
McKellop, H ;
Shen, FW ;
DiMaio, W ;
Lancaster, JG .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) :73-82
[5]
Development status of terumo implantable left ventricular assist system [J].
Nojiri, C ;
Kijima, T ;
Maekawa, J ;
Horiuchi, K ;
Kido, T ;
Sugiyama, T ;
Mori, T ;
Sugiura, N ;
Asada, T ;
Umemura, W ;
Ozaki, T ;
Suzuki, M ;
Akamatsu, T ;
Westaby, S ;
Katsumata, T ;
Saito, S .
ARTIFICIAL ORGANS, 2001, 25 (05) :411-413
[6]
Development of the Baylor Gyro permanently implantable centrifugal blood pump as a biventricular assist device [J].
Nonaka, K ;
Linneweber, J ;
Ichikawa, S ;
Yoshikawa, T ;
Kawahito, S ;
Mikami, M ;
Motomura, T ;
Ishitoya, H ;
Nishimura, I ;
Oestmann, D ;
Glueck, J ;
Schima, H ;
Wolner, E ;
Shinohara, T ;
Nosé, Y .
ARTIFICIAL ORGANS, 2001, 25 (09) :675-682
[7]
Development of a totally implantable biventricular bypass centrifugal blood pump system [J].
Nosé, Y ;
Nakata, K ;
Yoshikawa, M ;
Letsou, GV ;
Fujisawa, A ;
Wolner, E ;
Schima, H .
ANNALS OF THORACIC SURGERY, 1999, 68 (02) :775-779
[8]
VentrAssist hydrodynamically suspended, open, centrifugal blood pump [J].
Watterson, PA ;
Woodard, JC ;
Ramsden, VS ;
Reizes, JA .
ARTIFICIAL ORGANS, 2000, 24 (06) :475-477
[9]
Yamane T., 2002, J Artif Organs, V5, P149, DOI [10.1007/s100470200027, DOI 10.1007/S100470200027]