Primary stability in cementless femoral stems:: custom-made versus conventional femoral prosthesis

被引:93
作者
Götze, C
Steens, W
Vieth, V
Poremba, C
Claes, L
Steinbeck, J
机构
[1] Univ Hosp Muenster, Dept Orthopaed, D-48129 Munster, Germany
[2] Univ Munster, Dept Clin Radiol, D-4400 Munster, Germany
[3] Univ Munster, Dept Pathol, D-4400 Munster, Germany
[4] Univ Ulm, Inst Orthopaed Res & Biomech, D-89069 Ulm, Germany
关键词
total hip replacement; custom-made femoral stem; mechanical testing; initial stability; in vitro;
D O I
10.1016/S0268-0033(02)00012-8
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Objective. To compare a custom-made femoral stem with a conventional cementless stem for initial stability in human femurs. Design. The bone-femoral prosthesis interface motion was measured in vitro using displacement transducers. Combined axial and torsional loads simulating single-leg stance, loading were applied. Background. The custom-made prosthesis is a cementless femoral stem fabricated from the patient's CT- reconstruction. To justify its clinical use its design has to be tested regarding his primary stability. Methods. Seven pairs of human cadaveric femurs were used for testing the custom-made prosthesis versus a conventional cementless stem. Subsidence, rotation and interface motion were measured with load cycles up to 2000 N. Results. The critical interface motion of 150 pm for the cementless prosthesis was not exceed for both types of prosthesis. The custom-made prosthesis enlarged in its metaphyseal part showed a decrease of dynamic micromotions and rotation behavior in the proximal region compared to the Alloclassic prosthesis stem. Conclusions. The initial stability of the custom-made femoral stem was comparable to that of a conventional femoral stem.
引用
收藏
页码:267 / 273
页数:7
相关论文
共 28 条
[1]
Bachus KN, 1999, CLIN ORTHOP RELAT R, P248
[2]
BARGAR WL, 1989, CLIN ORTHOP RELAT R, V249, P73
[3]
HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[4]
Three-dimensional primary stability of cementless femoral stems [J].
Buhler, DW ;
Berlemann, U ;
Lippuner, K ;
Jaeger, P ;
Nolte, LP .
CLINICAL BIOMECHANICS, 1997, 12 (02) :75-86
[5]
MICROMOTION OF CEMENTED AND UNCEMENTED FEMORAL COMPONENTS [J].
BURKE, DW ;
OCONNOR, DO ;
ZALENSKI, EB ;
JASTY, M ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (01) :33-37
[6]
CAPELLO WN, 1989, CLIN ORTHOP RELAT R, V249, P56
[7]
CHARNLEY J, 1965, J Bone Joint Surg Br, V47, P56
[8]
Initial stability of fully and partially cemented femoral stems [J].
Claes, L ;
Fiedler, S ;
Ohnmacht, M ;
Duda, GN .
CLINICAL BIOMECHANICS, 2000, 15 (10) :750-755
[9]
Dai KR, 1985, ORTHOP T, V10, P99
[10]
TELEMETRIC FORCE MEASUREMENTS ACROSS THE HIP AFTER TOTAL ARTHROPLASTY [J].
DAVY, DT ;
KOTZAR, GM ;
BROWN, RH ;
HEIPLE, KG ;
GOLDBERG, VM ;
HEIPLE, KG ;
BERILLA, J ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (01) :45-50