Sample injection in capillary electrochromatography by heart-cut technique

被引:6
作者
Kahle, V
Kost'ál, V
Zeisbergerová, M
机构
[1] Acad Sci Czech Republ, Inst Analyt Chem, CS-61142 Brno, Czech Republic
[2] Palacky Univ, Fac Sci, Olomouc 77146, Czech Republic
[3] Masaryk Univ, Fac Sci, CS-61137 Brno, Czech Republic
关键词
electrochromatography; injection methods; sample handling; gradient elution; automation;
D O I
10.1016/j.chroma.2004.04.054
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The splitter working in heart-cut regime was used for sample injection in capillary electrochromatography. The principle was implemented in an automated microgradient system allowing to inject from microlitre down to nanolitre volumes with high repeatability and minimal extra-column band broadening. The apparatus is able to deliver discrete volumes of liquids at a preset volumetric flow rate and to stop and restore the flow at any moment. This brings a high degree of liquid manipulation flexibility. An extremely low split ratio is sufficient during the analysis, which saves mobile phase consumption substantially. The key parameters influencing the function of the heart-cut splitter were characterised. The function of the apparatus was demonstrated under isocratic, preconcentration and gradient capillary electrochromatography separation conditions. In all cases the statistic evaluation of the main parameters was performed, showing that high repeatability of retention times, peak heights and areas was achieved. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:259 / 265
页数:7
相关论文
共 35 条
[1]   Silica-based monoliths for capillary electrochromatography: Methods of fabrication and their applications in analytical separations [J].
Allen, D ;
El Rassi, Z .
ELECTROPHORESIS, 2003, 24 (22-23) :3962-3976
[2]   PRESSURIZED GRADIENT ELECTRO-HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
BEHNKE, B ;
BAYER, E .
JOURNAL OF CHROMATOGRAPHY A, 1994, 680 (01) :93-98
[3]  
Colón LA, 2000, ELECTROPHORESIS, V21, P3965, DOI 10.1002/1522-2683(200012)21:18<3965::AID-ELPS3965>3.0.CO
[4]  
2-T
[5]   ELECTRIC SAMPLE SPLITTER FOR CAPILLARY ZONE ELECTROPHORESIS [J].
DEML, M ;
FORET, F ;
BOCEK, P .
JOURNAL OF CHROMATOGRAPHY, 1985, 320 (01) :159-165
[6]  
Dermaux A, 1999, ELECTROPHORESIS, V20, P3027, DOI 10.1002/(SICI)1522-2683(19991001)20:15/16<3027::AID-ELPS3027>3.0.CO
[7]  
2-Z
[8]   Recent applications in capillary electrochromatography [J].
Eeltink, S ;
Rozing, GR ;
Kok, WT .
ELECTROPHORESIS, 2003, 24 (22-23) :3935-3961
[9]   Step-gradient capillary electrochromatography [J].
Euerby, MR ;
Gilligan, D ;
Johnson, CM ;
Bartle, KD .
ANALYST, 1997, 122 (10) :1087-1088
[10]   Polymeric monolithic stationary phases for capillary electrochromatography [J].
Hilder, EF ;
Svec, F ;
Fréchet, JMJ .
ELECTROPHORESIS, 2002, 23 (22-23) :3934-3953