Interference by cellular phones with permanent implanted pacemakers:: an update

被引:42
作者
Hekmat, K
Salemink, B
Lauterbach, G
Schwinger, RHG
Südkamp, M
Weber, HJ
Mehlhorn, U
机构
[1] Univ Cologne, Dept Cardiothorac Surg, D-50924 Cologne, Germany
[2] Univ Aachen, Dept Cardiovasc Engn Julich, Aachen, Germany
[3] Univ Cologne, Dept Cardiol, D-5000 Cologne 41, Germany
来源
EUROPACE | 2004年 / 6卷 / 04期
关键词
cellular telephones; interference; feedthru filter; pacemaker;
D O I
10.1016/j.eupc.2004.03.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims The aim of this study was to test pacemakers with feedthru fitters for interference with a digital cellular telephone. Methods The study comprised 100 patients having their first pacemaker implantation between January 2001 and May 2003. A GSM-standard cellular phone was tested in the standby, dialing and operating mode against 23 single-chamber and 77 dual-chamber pacemakers. Continuous surface electrocardiograms, intracardiac electrograms, and marker channels were recorded when calls were made by a land phone to the cellular phone. Results In 2 patients we observed pacing inhibition with the cellular phone positioned directly above the pacemaker pocket. The unipolar and bipolar ventricular sensitivity setting was 0.25 mV in one dual-chamber pacemaker, and when we changed the sensitivity to 0.50 mV and higher no interference could be detected. The second inhibition occurred in a single-chamber pacemaker with unipolar and bipolar ventricular sensitivity setting of 0.5 mV, where a sensitivity change to 1.0 mV eliminated the interference. Conclusion Anticipating a correct setting of ventricular sensitivity, currently available pacemakers equipped with feedthru fitters do not show any interference with cellular phones. Since interference was only observed with sensitivity settings below 0.50 mV, we recommend that permanent programming of ventricular sensitivity should be set at 2.0 mV and higher. (C) 2004 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:363 / 369
页数:7
相关论文
共 20 条
[1]  
Altamura G, 1997, EUR HEART J, V18, P1632
[2]   DO EUROPEAN GSM MOBILE CELLULAR PHONES POSE A POTENTIAL RISK TO PACEMAKER PATIENTS [J].
BARBARO, V ;
BARTOLINI, P ;
DONATO, A ;
MILITELLO, C ;
ALTAMURA, G ;
AMMIRATI, F ;
SANTINI, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (06) :1218-1224
[3]   Electromagnetic interference of analog cellular telephones with pacemakers [J].
Barbaro, V ;
Bartolini, P ;
Donato, A ;
Militello, C .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1410-1418
[4]   Interference of cellular phones with implanted permanent pacemakers [J].
Chen, WH ;
Lau, CP ;
Leung, SK ;
Ho, DSW ;
Lee, ISF .
CLINICAL CARDIOLOGY, 1996, 19 (11) :881-886
[5]   Influence of D-net (EUROPEAN GSM-standard) cellular telephones on implanted pacemakers in children [J].
Elshershari, H ;
Çeliker, A ;
Özer, S ;
Özme, S .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (09) :1328-1330
[6]   Interference with cardiac pacemakers by cellular telephones [J].
Hayes, DL ;
Wang, PJ ;
Reynolds, DW ;
Estes, NAM ;
Griffith, JL ;
Steffens, RA ;
Carlo, GL ;
Findlay, GK ;
Johnson, CM .
NEW ENGLAND JOURNAL OF MEDICINE, 1997, 336 (21) :1473-1479
[7]   State of the science: Pacemaker and defibrillator interference from wireless communication devices [J].
Hayes, DL ;
Carrillo, RG ;
Findlay, GK ;
Embrey, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1419-1430
[8]   Do mobile cellular phones in the European C and D net pose a potential risk to patients with implanted cardiac pacemakers? [J].
Hofgartner, F ;
Muller, T ;
Sigel, H .
DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1996, 121 (20) :646-652
[9]   Electromagnetic interference of pacemakers by mobile phones [J].
Irnich, W ;
Batz, L ;
Muller, R ;
Tobisch, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10) :1431-1446
[10]   Electronic security systems and active implantable medical devices [J].
Irnich, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (08) :1235-1258