Fabrication of a versatile microanalytical system without need for clean room conditions

被引:19
作者
Szekely, L
Freitag, R [1 ]
机构
[1] Swiss Fed Inst Technol, Fac Basic Sci, Lab Chem Biotechnol, CH-1015 Lausanne, Switzerland
[2] Univ Bayreuth, Fac Engn Sci, Chair Proc Biotechnol, Bayreuth, Germany
关键词
microfluidic chip; meniscus; adhesion forces;
D O I
10.1016/j.aca.2004.02.016
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper, we address the problem of designing microanalytical systems that are easily interfaced to the general laboratory environment ("macro world") and that do not require any special facilities such as clean rooms for their fabrication. The separation chip is realised in glass, the channels in particular by sawing with a 70 mum resinoid blade. While this restricts us to the fabrication of straight channels, it is much faster and less complex than any other method used at present for channel production in microfluidic chips. In view of a future application in pressure or voltage driven nano-chromatography, a monolithic stationary phase was synthesised in the separation channel of the chip. A multifunctional connection and support unit containing optical probes, reservoirs, as well as high pressure and high voltage connectors, constituted the interface to the macro world. Depending on the procedure used to bond the upper and the lower part of the microfluidic chip (adhesion forces or high temperature bonding) chips inserted in the unit could withstand upper pressure limits of 130 and 150 bar, respectively without leakage or disintegration. The system was then characterised in terms of its fluid dynamics under pressure and voltage driven conditions and its functionality was demonstrated in a preliminary experiment. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:39 / 47
页数:9
相关论文
共 36 条
[1]   Planar quartz chips with submicron channels for two-dimensional capillary electrophoresis applications [J].
Becker, H ;
Lowack, K ;
Manz, A .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1998, 8 (01) :24-28
[2]   Microfabricated devices in biotechnology and biochemical processing [J].
Chován, T ;
Guttman, A .
TRENDS IN BIOTECHNOLOGY, 2002, 20 (03) :116-122
[3]   Microchip injection and separation anomalies due to pressure effects [J].
Crabtree, HJ ;
Cheong, ECS ;
Tilroe, DA ;
Backhouse, CJ .
ANALYTICAL CHEMISTRY, 2001, 73 (17) :4079-4086
[4]   Comparison of standard capillary and chip separations of sodium dodecylsulfate-protein complexes [J].
Deyl, Z ;
Miksík, I ;
Eckhardt, A .
JOURNAL OF CHROMATOGRAPHY A, 2003, 990 (1-2) :153-158
[5]   Integrated capillary electrophoresis on flexible silicone microdevices: Analysis of DNA restriction fragments and detection of single DNA molecules on microchips [J].
Effenhauser, CS ;
Bruin, GJM ;
Paulus, A ;
Ehrat, M .
ANALYTICAL CHEMISTRY, 1997, 69 (17) :3451-3457
[6]   Electroosmosis- and pressure-driven chromatography in chips using continuous beds [J].
Ericson, C ;
Holm, J ;
Ericson, T ;
Hjertén, S .
ANALYTICAL CHEMISTRY, 2000, 72 (01) :81-87
[7]   A high-throughput continuous sample introduction interface for microfluidic chip-based capillary electrophoresis systems [J].
Fang, Q ;
Xu, GM ;
Fang, ZL .
ANALYTICAL CHEMISTRY, 2002, 74 (06) :1223-1231
[8]   Characterisation of a micro-total analytical system for the determination of nitrite with spectrophotometric detection [J].
Greenway, GM ;
Haswell, SJ ;
Petsul, PH .
ANALYTICA CHIMICA ACTA, 1999, 387 (01) :1-10
[9]   Sample manipulation in micro total analytical systems [J].
Greenwood, PA ;
Greenway, GM .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2002, 21 (11) :726-740
[10]   CAPILLARY ELECTROPHORESIS AND SAMPLE INJECTION SYSTEMS INTEGRATED ON A PLANAR GLASS CHIP [J].
HARRISON, DJ ;
MANZ, A ;
FAN, ZH ;
LUDI, H ;
WIDMER, HM .
ANALYTICAL CHEMISTRY, 1992, 64 (17) :1926-1932