Improved method for impedance measurements of mammalian cells

被引:24
作者
Ruemenapp, Christine [1 ]
Remm, Margarete [2 ]
Wolf, Bernhard [1 ,2 ]
Gleich, Bernhard [1 ]
机构
[1] Tech Univ Munich, Zent Inst Med Tech, D-85748 Garching, Germany
[2] Tech Univ Munich, Heinz Nixdorf Lehrstuhl Med Elekt, D-80333 Munich, Germany
关键词
Impedance spectroscopy; Electric circuit; Interdigitated electrode structures; Mathematical model; L929; Triton X-100; ENDOTHELIAL-CELLS; SPECTROSCOPY; MONITOR; MICROELECTRODE; BIOSENSOR; SURFACES; BEHAVIOR;
D O I
10.1016/j.bios.2009.02.030
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This study describes the use of impedance spectroscopy as a non-invasive tool to monitor cell proliferation and morphology and to obtain better measurements of cell kinetics. The goal was to determine test frequencies at which the cell behaviour towards different chemicals and toxins can be monitored with good time resolution. For optimisation of the test frequency it is in the first instance necessary to measure an impedance spectrum of the confluent cell layer. Secondly, the frequency of the highest peak in the spectrum will be determined and further experiments will be carried out at this fixed test frequency. In contrast to measurements at the mainly used 10 kHz (33%), signal differences between a confluent cell layer and cells after treatment with 0.2% Triton X-100 of up to 60% were obtained. In this work the measurements were carried out using interdigitated electrode structures (IDES) on a glass substrate. Impedance spectra of confluent L929 and MDA cell layers were obtained with the Solartron response analyzer. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:2915 / 2919
页数:5
相关论文
共 18 条
[1]   Non-invasive and microinvasive electrical impedance spectra of skin cancer -: a comparison between two techniques [J].
Åberg, P ;
Geladi, P ;
Nicander, I ;
Hansson, J ;
Holmgren, U ;
Ollmar, S .
SKIN RESEARCH AND TECHNOLOGY, 2005, 11 (04) :281-286
[2]   Bioelectrical impedance assay to monitor changes in cell shape during apoptosis [J].
Arndt, S ;
Seebach, J ;
Psathaki, K ;
Galla, HJ ;
Wegener, J .
BIOSENSORS & BIOELECTRONICS, 2004, 19 (06) :583-594
[3]  
Barsoukov E, 2005, IMPEDANCE SPECTROSCOPY: THEORY, EXPERIMENT, AND APPLICATIONS, 2ND EDITION, pXII
[4]   Online monitoring of BALB/3T3 metabolism and adhesion with multiparametric chip-based system [J].
Ceriotti, L. ;
Kob, A. ;
Drechsler, S. ;
Ponti, J. ;
Thedinga, E. ;
Colpo, P. ;
Ehret, R. ;
Rossi, F. .
ANALYTICAL BIOCHEMISTRY, 2007, 371 (01) :92-104
[5]   Assessment of cytotoxicity by impedance spectroscopy [J].
Ceriotti, Laura ;
Ponti, Jessica ;
Colpo, Pascal ;
Sabbioni, Enrico ;
Rossi, Francois .
BIOSENSORS & BIOELECTRONICS, 2007, 22 (12) :3057-3063
[6]   Electrical characterization of human mesenchymal stem cell growth on microelectrode [J].
Cho, Sungbo ;
Thielecke, Hagen .
MICROELECTRONIC ENGINEERING, 2008, 85 (5-6) :1272-1274
[7]   Monitoring of cellular behaviour by impedance measurements on interdigitated electrode structures [J].
Ehret, R ;
Baumann, W ;
Brischwein, M ;
Schwinde, A ;
Stegbauer, K ;
Wolf, B .
BIOSENSORS & BIOELECTRONICS, 1997, 12 (01) :29-41
[8]   MONITORING FIBROBLAST BEHAVIOR IN TISSUE-CULTURE WITH AN APPLIED ELECTRIC-FIELD [J].
GIAEVER, I ;
KEESE, CR .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA-BIOLOGICAL SCIENCES, 1984, 81 (12) :3761-3764
[9]   A MORPHOLOGICAL BIOSENSOR FOR MAMMALIAN-CELLS [J].
GIAEVER, I ;
KEESE, CR .
NATURE, 1993, 366 (6455) :591-592
[10]   MICROMOTION OF MAMMALIAN-CELLS MEASURED ELECTRICALLY [J].
GIAEVER, I ;
KEESE, CR .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1991, 88 (17) :7896-7900