Complete loss of ICD programmability after magnetic resonance imaging

被引:44
作者
Fiek, M [1 ]
Remp, T [1 ]
Reithmann, C [1 ]
Steinbeck, G [1 ]
机构
[1] Univ Munich, Med Hosp 1, D-81377 Munich, Germany
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2004年 / 27卷 / 07期
关键词
implantable cardioverter defibrillator; magnetic resonance imaging;
D O I
10.1111/j.1540-8159.2004.00573.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this case report is to describe the effects of an MRI performed on a patient without realizing that an ICD has been previously implanted. After a few seconds of imaging the adversity was recognized and the examination was stopped immediately. The patient was not pacemaker dependent and had neither physical complaints nor electrocardiographic changes in the surface ECG. A consecutively performed ICD assessment showed a backup mode with standard parameters for pacing (VVI 50 beats/min) and arrhythmia detection and treatment. The device could not be programmed by the external programmer. With the exception of printing out the parameters, all software functions were no longer feasible. A device examination by the manufacturer after ICD replacement showed that a major portion of the device memory was corrupt. Even ICDs of a newer generation are susceptible to magnetic interference, with the danger of complete loss of programmability.
引用
收藏
页码:1002 / 1004
页数:3
相关论文
共 11 条
[1]   Effects of magnetic resonance Imaging on cardiac pacemakers and electrodes [J].
Achenbach, S ;
Moshage, W ;
Diem, B ;
Bieberle, T ;
Schibgilla, V ;
Bachmann, K .
AMERICAN HEART JOURNAL, 1997, 134 (03) :467-473
[2]   NUCLEAR MAGNETIC-RESONANCE IMAGING IN A PATIENT WITH A DDD PACEMAKER [J].
ALAGONA, P ;
TOOLE, JC ;
MANISCALCO, BS ;
GLOVER, MU ;
ABERNATHY, GT ;
PRIDA, XE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (04) :619-619
[3]   Implantable cardioverter defibrillator dysfunction during and after magnetic resonance imaging [J].
Anfinsen, OG ;
Berntsen, RF ;
Aass, H ;
Kongsgaard, E ;
Amlie, JP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (09) :1400-1402
[4]   Pacing in magnetic resonance imaging environment:: Clinical and technical considerations on compatibility [J].
Duru, F ;
Luechinger, R ;
Scheidegger, MB ;
Lüscher, TF ;
Boesiger, P ;
Candinas, R .
EUROPEAN HEART JOURNAL, 2001, 22 (02) :113-124
[5]   Safe performance of magnetic resonance imaging on five patients with permanent cardiac pacemakers [J].
Gimbel, JR ;
Johnson, D ;
Levine, PA ;
Wilkoff, BL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (06) :913-919
[6]   EFFECT OF 1.5 TESLA NUCLEAR-MAGNETIC-RESONANCE IMAGING SCANNER ON IMPLANTED PERMANENT PACEMAKERS [J].
HAYES, DL ;
HOLMES, DR ;
GRAY, JE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (04) :782-786
[7]   NUCLEAR-MAGNETIC-RESONANCE IMAGING IN A PATIENT WITH A PACEMAKER [J].
INBAR, S ;
LARSON, J ;
BURT, T ;
MAFEE, M ;
EZRI, MD .
AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 1993, 305 (03) :174-175
[8]  
JOHNSON D, 1994, PACE, V17, P772
[9]   EFFECTS OF NUCLEAR-MAGNETIC-RESONANCE IMAGING ON CARDIAC-PACEMAKERS [J].
LAUCK, G ;
VONSMEKAL, A ;
WOLKE, S ;
SEELOS, KC ;
JUNG, W ;
MANZ, M ;
LUDERITZ, B .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (08) :1549-1555
[10]   Force and torque effects of a 1.5-Tesla MRI scanner on cardiac pacemakers and ICDs [J].
Luechinger, R ;
Duru, F ;
Scheidegger, MB ;
Boesiger, P ;
Candinas, R .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (02) :199-205