Towards Optimized Acquisition Scheme for Multiprojection Correlation Imaging of Breast Cancer

被引:3
作者
Chawla, Amarpreet S. [1 ,2 ]
Saunders, Robert S. [1 ]
Singh, Swatee [1 ,2 ]
Lo, Joseph Y. [1 ,2 ,3 ]
Samei, Ehsan [1 ,2 ,3 ,4 ]
机构
[1] Duke Univ, Duke Adv Imaging Labs, Dept Radiol, Durham, NC 27705 USA
[2] Duke Univ, Duke Adv Imaging Labs, Dept Biomed Engn, Durham, NC 27705 USA
[3] Duke Univ, Duke Adv Imaging Labs, Dept Med Phys, Durham, NC 27705 USA
[4] Duke Univ, Duke Adv Imaging Labs, Dept Phys, Durham, NC 27705 USA
关键词
Multiprojection imaging; correlation imaging; breast tomosynthesis; FROC; CADe; TOMOSYNTHESIS; NODULES; NOISE;
D O I
10.1016/j.acra.2008.09.013
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. Correlation imaging (CI) is form of multiprojection imaging in which multiple images of a patient are acquired from slightly different angles. Information from these images is combined to make the final diagnosis. A critical factor affecting the performance of Cl is its data acquisition scheme. because nonoptimized acquisition may distort pathologic indicators. The authors describe a computer-aided detection (CADe) methodology to optimize the acquisition scheme of Cl for superior diagnostic accuracy. Materials and Methods. Images from 106 subjects were used. For each subject, 25 angular projections of a single breast were acquired. Projection images were supplemented with a simulated 3-mm three-dimensional lesion. Each projection was then processed using a traditional CADe algorithm at high sensitivity, followed by the reduction of false-positives by combining the geometric correlation information available from the multiple images. The performance of the CI system was determined in terms of free-response receiver-operating characteristic curves and the areas under receiver-operating characteristic Curves. For optimization, the components of acquisition, such as the number of projections and their angular span, were systematically changed to investigate which of the many possible combinations maximized the obtainable CADe sensitivity and specificity. Results. The performance of the CI system was improved by increasing the angular span. Increasing the number of angular projections beyond a certain number did not improve performance. Maximum performance was obtained between 7 and 10 projections spanning a maximum angular arc of 45 degrees. Conclusion. The findings suggest the existence of an optimum acquisition scheme for CI of the breast. CADe results confirmed earlier predictions on the basis of observer models. An optimized CI system may be an important diagnostic tool for improved breast cancer detection.
引用
收藏
页码:456 / 463
页数:8
相关论文
共 15 条
[1]   Digital breast tomosynthesis using an amorphous selenium flat panel detector [J].
Bissonnette, M ;
Hansroul, M ;
Masson, E ;
Savard, S ;
Cadieux, S ;
Warmoes, P ;
Gravel, D ;
Agopyan, J ;
Polischuk, B ;
Haerer, W ;
Mertelmeier, T ;
Lo, JY ;
Chen, Y ;
Dobbins, JT ;
Jesneck, JL ;
Singh, S .
Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 :529-540
[2]   Characterization of difference of Gaussian filters in the detection of mammographic regions [J].
Catarious, David M., Jr. ;
Baydush, Alan H. ;
Floyd, Carey E., Jr. .
MEDICAL PHYSICS, 2006, 33 (11) :4104-4114
[3]  
Chan H.-P., 2007, PROC SPIE, V6514, P1
[4]  
CHAWLA A, 2008, P INT WORKSH DIG MAM, V9, P635
[5]  
Chawla A., 2007, PROC SPIE, V6510, P1
[6]   A mathematical model platform for optimizing a multiprojection breast imaging system [J].
Chawla, Amarpreet S. ;
Samei, Ehsan ;
Saunders, Robert S. ;
Lo, Joseph Y. ;
Baker, Jay A. .
MEDICAL PHYSICS, 2008, 35 (04) :1337-1345
[7]   Digital x-ray tomosynthesis: current state of the art and clinical potential [J].
Dobbins, JT ;
Godfrey, DJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (19) :R65-R106
[8]   Optimized approach to decision fusion of heterogeneous data for breast cancer diagnosis [J].
Jesneck, Jonathan L. ;
Nolte, Loren W. ;
Baker, Jay A. ;
Floyd, Carey E. ;
Lo, Joseph Y. .
MEDICAL PHYSICS, 2006, 33 (08) :2945-2954
[9]   Effects of quantum noise and binocular summation on dose requirements in stereoradiography [J].
Maidment, ADA ;
Bakic, PR ;
Albert, M .
MEDICAL PHYSICS, 2003, 30 (12) :3061-3071
[10]   Computerized mass detection for digital breast tomosynthesis directly from the projection images [J].
Reiser, I ;
Nishikawa, RM ;
Giger, ML ;
Wu, T ;
Rafferty, EA ;
Moore, R ;
Kopans, DB .
MEDICAL PHYSICS, 2006, 33 (02) :482-491