Development and characterization of microfabricated disposable gold working electrodes for high-performance ion chromatography and integrated pulsed amperometric detection

被引:32
作者
Cheng, J [1 ]
Jandik, P [1 ]
Avdalovic, N [1 ]
机构
[1] Dionex Chem Corp, Sunnyvale, CA 94088 USA
关键词
D O I
10.1021/ac0204754
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We have developed a new type of microfabricated thin-film electrode on polymeric substrates. The microfabrication process allows for inexpensive and reproducible mass production of disposable working electrodes for high-performance ion chromatography and integrated pulsed amperometric detection (IPAD). These microfabricated electrodes are disposable and have been optimized for use in flow-through low-dead-volume electrochemical cells. The analytical performance of microfabricated gold electrodes was characterized with the help of the IPAD method for amino acid detection under alkaline conditions required for anion-exchange separations. When used with a new optimized six-potential IPAD waveform, the electrodes functioned properly for weeks. Compared to nondisposable working electrodes, the disposable working electrodes generated equal or better results in the limit of detection, linearity of calibration, and reproducibility. Disposable electrodes make it possible to avoid polishing and reconditioning, which are required with nondisposable electrodes.
引用
收藏
页码:572 / 579
页数:8
相关论文
共 26 条
[21]   Improved long-term reproducibility for pulsed amperometric detection of carbohydrates via a new quadruple-potential waveform [J].
Rocklin, RD ;
Clarke, AP ;
Weitzhandler, M .
ANALYTICAL CHEMISTRY, 1998, 70 (08) :1496-1501
[22]   SELF-ASSEMBLED OMEGA-HYDROXYALKANETHIOL MONOLAYERS WITH INTERNAL FUNCTIONALITIES - ELECTROCHEMICAL AND INFRARED STRUCTURAL CHARACTERIZATIONS OF ETHER-CONTAINING MONOLAYERS [J].
SINNIAH, K ;
CHENG, J ;
TERRETTAZ, S ;
REUTTROBEY, JE ;
MILLER, CJ .
JOURNAL OF PHYSICAL CHEMISTRY, 1995, 99 (39) :14500-14505
[23]   BEHAVIOR OF AN ELECTROCHEMICAL DETECTOR USED IN LIQUID-CHROMATOGRAPHY AND CONTINUOUS-FLOW VOLTAMMETRY .1. MASS TRANSPORT LIMITED CURRENT [J].
WEBER, SG ;
PURDY, WC .
ANALYTICA CHIMICA ACTA, 1978, 100 (SEP) :531-544
[24]   Eliminating amino acid and peptide interference in high-performance anion-exchange pulsed amperometric detection glycoprotein monosaccharide analysis [J].
Weitzhandler, M ;
Pohl, C ;
Rohrer, J ;
Narayanan, L ;
Slingsby, R ;
Avdalovic, N .
ANALYTICAL BIOCHEMISTRY, 1996, 241 (01) :128-134
[25]   COMPARISON OF PULSED COULOMETRIC DETECTION AND POTENTIAL-SWEEP PULSED COULOMETRIC DETECTION FOR UNDERIVATIZED AMINO-ACIDS IN LIQUID-CHROMATOGRAPHY [J].
WELCH, LE ;
LACOURSE, WR ;
MEAD, DA ;
JOHNSON, DC ;
HU, T .
ANALYTICAL CHEMISTRY, 1989, 61 (06) :555-559
[26]   THE ELECTROCHEMICAL DESORPTION OF N-ALKANETHIOL MONOLAYERS FROM POLYCRYSTALLINE AU AND AG ELECTRODES [J].
WIDRIG, CA ;
CHUNG, C ;
PORTER, MD .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1991, 310 (1-2) :335-359