A geometric calibration method for cone beam CT systems

被引:180
作者
Yang, Kai
Kwan, Alexander L. C.
Miller, DeWitt F.
Boone, John M.
机构
[1] Univ Calif Davis, Med Ctr, Dept Radiol, Sacramento, CA 95817 USA
[2] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
关键词
cone-beam; CT; geometric calibration;
D O I
10.1118/1.2198187
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Cone beam CT systems are being deployed in large numbers for small animal imaging, dental imaging, and other specialty applications. A new high-precision inethod for cone beam CT system calibration is presented in this paper. It uses multiple projection images acquired from rotating point-like objects (metal ball bearings) and the angle information generated from the rotating gantry system is also used. It is assumed that the whole system has a mechanically stable rotation center and that the detector does not have severe out-of-plane rotation (<2 degrees). Simple geometrical relationships between the orbital paths of individual BBs and five system parameters were derived. Computer simulations were employed to validate the accuracy of this method in the presence of noise. Equal or higher accuracy was achieved compared with previous methods. This method was implemented for the geometrical calibration of both a micro CT scanner and a breast CT scanner. The reconstructed tomographic images demonstrated that the proposed method is robust and easy to implement with high precision. (C) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1695 / 1706
页数:12
相关论文
共 10 条
[1]   Accurate technique for complete geometric calibration of cone-beam computed tomography systems [J].
Cho, YB ;
Moseley, DJ ;
Siewerdsen, JH ;
Jaffray, DA .
MEDICAL PHYSICS, 2005, 32 (04) :968-983
[2]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[3]   ESTIMATION OF GEOMETRICAL PARAMETERS FOR FAN BEAM TOMOGRAPHY [J].
GULLBERG, GT ;
TSUI, BMW ;
CRAWFORD, CR ;
EDGERTON, ER .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (12) :1581-1594
[4]   ESTIMATION OF GEOMETRICAL PARAMETERS AND COLLIMATOR EVALUATION FOR CONE BEAM TOMOGRAPHY [J].
GULLBERG, GT ;
TSUI, BMW ;
CRAWFORD, CR ;
BALLARD, JG ;
HAGIUS, JT .
MEDICAL PHYSICS, 1990, 17 (02) :264-272
[5]   Three-dimensional artifact induced by projection weighting and misalignment [J].
Hsieh, J .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (04) :364-368
[6]  
Kak A.C. Slaney M., 1999, PRINCIPLES COMPUTERI
[7]   Implementation of a cone-beam reconstruction algorithm for the single-circle source orbit with embedded misalignment correction using homogeneous coordinates [J].
Karolczak, M ;
Schaller, S ;
Engelke, K ;
Lutz, A ;
Taubenreuther, U ;
Wiesent, K ;
Kalender, W .
MEDICAL PHYSICS, 2001, 28 (10) :2050-2069
[8]   Analytic method based on identification of ellipse parameters for scanner calibration in cone-beam tomography [J].
Noo, F ;
Clackdoyle, R ;
Mennessier, C ;
White, TA ;
Roney, TJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3489-3508
[9]   Alignment of a volumetric tomography system [J].
Stevens, GM ;
Saunders, R ;
Pelc, NJ .
MEDICAL PHYSICS, 2001, 28 (07) :1472-1481
[10]   Geometric misalignment and calibration in cone-beam tomography [J].
von Smekal, L ;
Kachelriess, M ;
Stepina, E ;
Kalender, WA .
MEDICAL PHYSICS, 2004, 31 (12) :3242-3266