Suppressing the temperature effect in near infrared spectroscopy by using orthogonal signal correction

被引:5
作者
Blanco, Marcelo [1 ]
Valdes, Damarih [1 ]
机构
[1] Univ Autonoma Barcelona, Dept Chem, Bellaterra 08193, Barcelona, Spain
关键词
near infrared spectroscopy; spectral treatments; multivariate calibration; temperature effect; orthogonal signal correction;
D O I
10.1255/jnirs.610
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
Temperature changes alter the near infrared spectra for liquid samples and substantially diminish the predictive ability of calibration models constructed from spectra that are not recorded under strictly controlled temperature conditions. The effect is especially significant in compounds forming hydrogen bonds, which exhibit shifts and intensity variations in their spectral absorption maxima and provide non-linear relationships between variables that require more complex models. In this work, we used the variable filtering method known as orthogonal signal correction (OSC) with a view to eliminate the spectral variability introduced by temperature change that is unrelated to the concentrations of the ingredients of an esterification reaction. Once the spurious spectral information was removed by the OSC algorithm, calibration models were constructed from spectra recorded over a broad temperature range in order to determine each individual component (PLS1) and the four reaction ingredients simultaneously (PLS2) in liquid samples. The OSC spectral treatment was found to provide more simple and accurate (RSEP < 5%) predictive models that those obtained in its absence.
引用
收藏
页码:155 / 160
页数:6
相关论文
共 25 条
[1]   Determination of alcohol content in beverages using short-wave near-infrared spectroscopy and temperature correction by transfer calibration procedures [J].
Barboza, FD ;
Poppi, RJ .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2003, 377 (04) :695-701
[2]  
Beebe K. R., 1998, Chemometrics: A Practical Guide, P26
[3]   Orthogonal signal correction in near infrared calibration [J].
Blanco, M ;
Coello, J ;
Montoliu, I ;
Romero, MA .
ANALYTICA CHIMICA ACTA, 2001, 434 (01) :125-132
[4]   Influence of temperature on the predictive ability of near infrared spectroscopy models [J].
Blanco, M ;
Valdés, D .
JOURNAL OF NEAR INFRARED SPECTROSCOPY, 2004, 12 (02) :121-126
[5]   Near-infrared spectra of Penicillium camemberti strains separated by extended multiplicative signal correction improved prediction of physical and chemical variations [J].
Decker, M ;
Nielsen, PV ;
Martens, H .
APPLIED SPECTROSCOPY, 2005, 59 (01) :56-68
[6]   Examination of spectral pretreatments for partial least-squares calibrations for chemical and physical properties of wheat [J].
Delwiche, SR ;
Graybosch, RA .
APPLIED SPECTROSCOPY, 2003, 57 (12) :1517-1527
[7]   Removal of interference signals due to water from in vivo near-infrared (NIR) spectra of mood glucose by region orthogonal signal correction (ROSC) [J].
Du, YP ;
Liang, YZ ;
Kasemsumran, S ;
Maruo, K ;
Ozaki, Y .
ANALYTICAL SCIENCES, 2004, 20 (09) :1339-1345
[8]   On orthogonal signal correction [J].
Fearn, T .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2000, 50 (01) :47-52
[9]   Piecewise orthogonal signal correction [J].
Feudale, RN ;
Tan, HW ;
Brown, SD .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2002, 63 (02) :129-138
[10]   Temperature robust multivariate calibration: an overview of methods for dealing with temperature influences on near infrared spectra [J].
Hageman, JA ;
Westerhuis, JA ;
Smilde, AK .
JOURNAL OF NEAR INFRARED SPECTROSCOPY, 2005, 13 (02) :53-62