RF heating due to conductive wires during MRI depends on the phase distribution of the transmit field

被引:87
作者
Yeung, CJ
Susil, RC
Atalar, E
机构
[1] Johns Hopkins Univ, Sch Med, Dept Biomed Engn, Baltimore, MD 21205 USA
[2] Johns Hopkins Univ, Sch Med, Dept Radiol, Baltimore, MD 21205 USA
关键词
safety; RF heating; interventional MRI; method of moments; metallic implants; guidewire;
D O I
10.1002/mrm.10310
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In many studies concerning wire heating during MR imaging, a "resonant wire length" that maximizes RF heating is determined. This may lead to the nonintuitive conclusion that adding more wire, so as to avoid this resonant length, will actually improve heating safety. Through a theoretical analysis using the method of moments, we show that this behavior depends on the phase distribution of the RF transmit field. If the RF transmit field has linear phase, with slope equal to the real part of the wavenumber in the tissue, long wires always heat more than short wires. In order to characterize the intrinsic safety of a device without reference to a specific body coil design, this maximum-tip heating phase distribution must be considered. Finally, adjusting the phase distribution of the electric field generated by an RF transmit coil may lead to an "implant-friendly" coil design.
引用
收藏
页码:1096 / 1098
页数:3
相关论文
共 11 条
[1]   A galerkin moment method for the analysis of an insulated antenna in a dissipative dielectric medium [J].
Atlamazoglou, PE ;
Uzunoglu, NK .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 1998, 46 (07) :988-996
[2]   THERAPEUTIC APPLICATIONS OF ELECTROMAGNETIC POWER [J].
GUY, AW ;
LEHMANN, JF ;
STONEBRIDGE, JB .
PROCEEDINGS OF THE IEEE, 1974, 62 (01) :55-75
[3]  
Harrington R. F., 1968, FIELD COMPUTATION MO
[4]   Active MR visualization of a vascular guidewire in vivo [J].
Ladd, ME ;
Zimmermann, GG ;
Quick, HH ;
Debatin, JF ;
Boesiger, P ;
von Schulthess, GK ;
McKinnon, GC .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :220-225
[5]  
Nitz WR, 2001, J MAGN RESON IMAGING, V13, P105, DOI 10.1002/1522-2586(200101)13:1<105::AID-JMRI1016>3.0.CO
[6]  
2-0
[7]  
POLK C, 1996, HDB BIOL EFFECTS ELE, P297
[8]  
Ramo S., 1994, FIELDS WAVES COMMUNI, P149
[9]   MR-guided percutaneous angioplasty: Assessment of tracking safety, catheter handling and functionality [J].
Wildermuth, S ;
Dumoulin, CL ;
Pfammatter, T ;
Maier, SE ;
Hofmann, E ;
Debatin, JF .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1998, 21 (05) :404-410
[10]   RF safety of wires in interventional MRI: Using a safety index [J].
Yeung, CJ ;
Susil, RC ;
Atalar, E .