Voltammetric and reference microelectrodes with integrated microchannels for flow through microvoltammetry. 1. The microcell

被引:22
作者
Keller, OC [1 ]
Buffle, J [1 ]
机构
[1] Univ Geneva, Dept Inorgan Analyt & Appl Chem, CABE, CH-1211 Geneva 4, Switzerland
关键词
D O I
10.1021/ac9905064
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper describes the construction of 2 microsensor units for on-line voltammetric detection inside a cylindrical microcell (a working microsensor unit and a reference and auxiliary microsensor unit), for application to heavy-metal analysis in complex media such as natural waters. Both microsensor units include a channel for the solution renewal in the microcell after analysis. The working microsensor, a Hg-plated Ir microelectrode, is protected against fouling with an agarose gel including a hydrophobic chromatographic phase (C18). The fabrication steps and the quality tests related to long-term use and reliability, as well as to precision, are described. The application of the protective gel layer against fouling by hydrophobic or surface active small molecules is of general application, reliable, and very efficient. The reference and auxiliary unit is composed by an iridium oxide based mini reference electrode, an auxiliary Pt electrode, and a circulation channel. It is built to enable its use inside a 700-mu m-diameter tubing connected to a hollow fiber supported liquid membrane lumen (volume, 5-10 mu L) for heavy-metal analysis. However, it can be used in any other microanalytical system. The reference electrode is sufficiently stable for voltammetric applications (1-2 mV drift/day), and its lifetime is more than one year. The Ti/IrO2 core is immersed in a pH-buffered agarose gel, to guarantee potential stability even when the electrode is immersed in variable pH solutions.
引用
收藏
页码:936 / 942
页数:7
相关论文
共 44 条
[11]  
DEROOIJ NF, 1981, P INT C KARG BAS, P156
[12]   PREPARATION AND PROPERTIES OF A MERCURY-PLATED IRIDIUM-BASED MICROELECTRODE [J].
DEVITRE, RR ;
TERCIER, ML ;
TSACOPOULOS, M ;
BUFFLE, J .
ANALYTICA CHIMICA ACTA, 1991, 249 (02) :419-425
[13]   EVALUATION OF A NEW SOLID-STATE REFERENCE ELECTRODE JUNCTION MATERIAL FOR ION-SELECTIVE ELECTRODES [J].
DIAMOND, D ;
MCENROE, E ;
MCCARRICK, M ;
LEWENSTAM, A .
ELECTROANALYSIS, 1994, 6 (11-12) :962-971
[14]   RUTHENIUM DIOXIDE - NEW ELECTRODE MATERIAL .1. BEHAVIOR IN ACID SOLUTIONS OF INERT ELECTROLYTES [J].
GALIZZIOLI, D ;
TANTARDINI, F ;
TRASATTI, S .
JOURNAL OF APPLIED ELECTROCHEMISTRY, 1974, 4 (01) :57-67
[15]   METAL-METAL OXIDE AND METAL-OXIDE ELECTRODES AS PH SENSORS [J].
GLAB, S ;
HULANICKI, A ;
EDWALL, G ;
INGMAN, F .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1989, 21 (01) :29-47
[16]  
GOSSNER K, 1981, J ELECTROANAL CHEM, V125, P347, DOI 10.1016/S0022-0728(81)80353-1
[17]   ELECTROCHROMISM IN ANODIC IRIDIUM OXIDE-FILMS .2. PH EFFECTS ON CORROSION STABILITY AND THE MECHANISM OF COLORATION AND BLEACHING [J].
GOTTESFELD, S ;
MCINTYRE, JDE .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1979, 126 (05) :742-750
[18]   QUANTITATIVE DESCRIPTION OF MACROCYCLE-MEDIATED CATION SEPARATIONS IN LIQUID MEMBRANES AND ON SILICA-GEL AS A FUNCTION OF SYSTEM PARAMETERS [J].
IZATT, RM ;
BRUENING, RL ;
BRADSHAW, JS ;
LAMB, JD ;
CHRISTENSEN, JJ .
PURE AND APPLIED CHEMISTRY, 1988, 60 (04) :453-460
[19]   PH-SENSITIVE SPUTTERED IRIDIUM OXIDE-FILMS [J].
KATSUBE, T ;
LAUKS, I ;
ZEMEL, JN .
SENSORS AND ACTUATORS, 1982, 2 (04) :399-410
[20]  
KELLER OC, 1997, THESIS U GENEVA