Correction for scatter in 3D brain PET using a dual energy window method

被引:79
作者
Grootoonk, S
Spinks, TJ
Sashin, D
Spyrou, NM
Jones, T
机构
[1] HAMMERSMITH HOSP,MRC,CYCLOTRON UNIT,CLIN SERV CTR,LONDON W12 0NN,ENGLAND
[2] UPMC,PET FAC,PITTSBURGH,PA
[3] UNIV SURREY,DEPT PHYS,GUILDFORD GU2 5XH,SURREY,ENGLAND
关键词
D O I
10.1088/0031-9155/41/12/013
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A method for scatter correction using dual energy window acquisition has been developed and implemented on data collected with a brain-PET tomograph operated in the septa retracted, 3D mode. Coincidence events are assigned to (i) an upper energy window where both photons deposit energy between 380 keV and 850 keV or (ii) a lower energy window where one or both photons deposit within 200 keV and 380 keV. Scaling parameters are derived from measurements of the ratios of counts from line sources due to scattered and unscattered events in the two energy windows in head-sized phantoms. A scaled subtraction of the two energy windows produces a distribution of scatter which is smoothed prior to subtraction from the upper energy window. In phantoms, the correction was found to restore the uniformity, contrast and linearity of activity concentration. Relative activity concentrations were restored to within 7% of their true values in a multicompartment phantom. The method was found to provide accurate correction for scattered events arising from activity outside the direct detector field of view. In a three-compartment phantom containing water, F-18 and C-11 scanned in dynamic, multiframe mode, the half-lives of the two isotopes were restored to within 2% of their true value.
引用
收藏
页码:2757 / 2774
页数:18
相关论文
共 28 条
[1]  
AXELSSON B, 1984, J NUCL MED, V25, P490
[2]   A CONVOLUTION-SUBTRACTION SCATTER CORRECTION METHOD FOR 3D PET [J].
BAILEY, DL ;
MEIKLE, SR .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (03) :411-424
[3]  
BAILEY DL, 1995, P IEEE MED IM C SAN, V2, P1046
[4]   ADDITION OF NOISE BY SCATTER CORRECTION METHODS IN PVI [J].
BARNEY, JS ;
HARROP, R ;
ATKINS, MS .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1551-1555
[5]  
BENDRIEM B, 1993, P IEEE MED IM C SAN, V3, P1779
[6]   CORRECTION FOR SCATTERED RADIATION IN A RING DETECTOR POSITRON CAMERA BY INTEGRAL TRANSFORMATION OF THE PROJECTIONS [J].
BERGSTROM, M ;
ERIKSSON, L ;
BOHM, C ;
BLOMQVIST, G ;
LITTON, J .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (01) :42-50
[7]   A MULTICRYSTAL 2-DIMENSIONAL BGO DETECTOR SYSTEM FOR POSITRON EMISSION TOMOGRAPHY [J].
CASEY, ME ;
NUTT, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :460-463
[8]  
CHERRY SR, 1993, J NUCL MED, V34, P671
[9]   A NORMALIZATION TECHNIQUE FOR 3D PET DATA [J].
DEFRISE, M ;
TOWNSEND, DW ;
BAILEY, D ;
GEISSBUHLER, A ;
MICHEL, C ;
JONES, T .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) :939-952
[10]  
Diffrient N., 1974, HUMANSCALE 1 2 3