Determination of bromide, bromate and other anions with ion chromatography and an inductively coupled plasma mass spectrometer as element-specific detector

被引:40
作者
Divjak, B
Novic, M
Goessler, W
机构
[1] Graz Univ, Inst Analyt Chem, A-8010 Graz, Austria
[2] Natl Inst Chem, SI-1001 Ljubljana, Slovenia
关键词
anions; bromate; bromide; oxyhalogens; halogens;
D O I
10.1016/S0021-9673(99)00896-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
An implementation of the Dionex IonPac AS12A analytical column with an element-specific ICP-MS detection is described for the simultaneous determination of halogen and oxyhalogen anions, sulfate, phosphate, selenite, selenate and arsenate. The chromatographic separation was achieved in less than 4 min with an aqueous 11 mM (NH4)(2)CO3 (pH 11.2, adjusted with aqueous ammonia) as eluent. Special emphasis was given to optimize the ICP-MS detection conditions for the reliable detection (RSD<50%) of bromate and bromide at a bromine concentration level of 1.0 mu g l(-1) with 50 mu l sample injection volume. In order to achieve the highest detector response for bromine species an ultrasonic nebulizer equipped with a membrane desolvator had to be employed. The detection limits (S/N=3, sample injection volume 50 mu l) obtained with the IC-ICP-MS after the optimization were 0.67 mu g l(-1) for BrO3-, 0.47 mu g l(-1) for Br-, 69 mu g l(-1) for ClO2-, 4 mu g l(-1) for Cl- 47 mu g l(-1) for ClO3-, 13 mu g l(-1) for SO42-, 36 mu g l(-1) for PO34-, 0.4 mu g l(-1) for SeO32-, 0.3 mu g l(-1) for SeO42-, and 0.4 mu g l(-1) for AsO43-. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:39 / 47
页数:9
相关论文
共 26 条
[1]   SIGNAL ENHANCEMENT OF ELEMENTS DUE TO THE PRESENCE OF CARBON-CONTAINING COMPOUNDS IN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
ALLAIN, P ;
JAUNAULT, L ;
MAURAS, Y ;
MERMET, JM ;
DELAPORTE, T .
ANALYTICAL CHEMISTRY, 1991, 63 (14) :1497-1498
[2]  
*COMM EUR COMM, 1994, PROP COUNC DIR CONC
[3]   Isotope dilution analysis of bromate in drinking water matrixes by ion chromatography with inductively coupled plasma mass spectrometric detection [J].
Creed, JT ;
Brockhoff, CA .
ANALYTICAL CHEMISTRY, 1999, 71 (03) :722-726
[4]   Determination of bromate in drinking waters by ion chromatography with inductively coupled plasma mass spectrometric detection [J].
Creed, JT ;
Magnuson, ML ;
Pfaff, JD ;
Brockhoff, C .
JOURNAL OF CHROMATOGRAPHY A, 1996, 753 (02) :261-267
[5]   Determination of bromate in the presence of brominated haloacetic acids by ion chromatography with inductively coupled plasma mass spectrometric detection [J].
Creed, JT ;
Magnuson, ML ;
Brockhoff, CA .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1997, 31 (07) :2059-2063
[6]   Bromide/bromate speciation by NTI-IDMS and ICP-MS coupled with ion exchange chromatography [J].
Diemer, J ;
Heumann, KG .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1997, 357 (01) :74-79
[7]  
*DION CORP, 1994, INST INSTR TROUBL GU
[8]   OZONATION OF BROMIDE-CONTAINING WATERS - KINETICS OF FORMATION OF HYPOBROMOUS ACID AND BROMATE [J].
HAAG, WR ;
HOIGNE, J .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1983, 17 (05) :261-267
[9]   ION CHROMATOGRAPHY OF INORGANIC ANIONS WITH POTENTIOMETRIC DETECTION USING A METALLIC COPPER ELECTRODE [J].
HADDAD, PR ;
ALEXANDER, PW ;
TROJANOWICZ, M .
JOURNAL OF CHROMATOGRAPHY, 1985, 321 (02) :363-374
[10]   ANALYSIS OF OXYHALIDE DISINFECTION BY-PRODUCTS AND OTHER ANIONS OF INTEREST IN DRINKING-WATER BY ION CHROMATOGRAPHY [J].
HAUTMAN, DP ;
BOLYARD, M .