共 7 条
Disposable thermo-pneumatic micropurnp for bio lab-on-a-chip application
被引:36
作者:
Ha, Seung-Mo
[2
]
Cho, Woong
[2
]
Ahn, Yoomin
[1
]
机构:
[1] Hanyang Univ, Dept Mech Engn, Ansan 426791, Gyeonggi, South Korea
[2] Hanyang Univ, Dept Mech Engn, Grad Sch, Seoul 133791, South Korea
关键词:
Thermo-pneumatic micropump;
Microscale check valve;
Detachable microfluidic device;
PDMS-glass hybrid chip;
SYSTEMS;
D O I:
10.1016/j.mee.2008.12.046
中图分类号:
TM [电工技术];
TN [电子技术、通信技术];
学科分类号:
0808 ;
0809 ;
摘要:
We present a disposable thermo-pneumatic micropump with microscale check valves for easy incorporation in a fully integrated biochip. The microfluidic device is comprised of biocompatible materials: glass and polydimethylsiloxane (PDMS). The chip consists of a glass substrate and three PDMS layers. in the chip, a Cr/Au heater is fabricated by micromachining on a glass substrate, while the flow channel, pump chamber, and valves are made by replica molding using layers of PDMS. The glass-based heater chip and the PDMS-based micropump chip are assembled and clamped together to enable operation of the system, and then the components are separable such that the heater chip can be used repeatedly and the PDMS chip can be used in a disposable manner. The operation of the micropump was investigated experimentally, as well as through simulations using fluid dynamics software. Superior pumping performance was achieved when the duty ratio of the voltage applied to the heater was 33%. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:1337 / 1339
页数:3
相关论文