A non-invasive patient registration and reference system for interactive intraoperative localization in intranasal sinus surgery

被引:11
作者
Hauser, R
Westermann, B
Probst, R
机构
[1] Department of Otorhinolaryngology, University of Basel, Basel
关键词
non-invasive referencing; optical localizing system; image guided surgery; computer assisted surgery; instrumentation; intranasal sinus surgery;
D O I
10.1243/0954411971534458
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A basic problem common to all systems for computer assisted surgery (CAS) is patient referencing, or the transfer of preoperative image data to the intraoperative pathology. The authors describe a highly precise CAS system with non-invasive referencing that can be used in ear, nose and throat (ENT) surgery of the paranasal sinuses. It is based on optical digitizing with several custom-made self-localizing surgical instruments. The accuracy of the system was tested in an experimental model using a plastic head. Measurements of repositioning the reference bow had a mean error of 0.81 mm+/-0.31 mm. The system was evaluated clinically with 11 patients who received surgery for different pathologies of the paranasal sinuses. These trials met with a high rate of success and specific results are reported.
引用
收藏
页码:327 / 334
页数:8
相关论文
共 11 条
[1]   A NEW MODEL FOR NONINVASIVE, REPRODUCIBLE FIXATION OF A STEREOTAXIC FRAME USING AN ORTHODONTIC RESIN PLATE [J].
CARINI, S ;
CALCAGNO, E ;
TORTORIDONATI, P ;
ANDREUSSI, L .
ACTA NEUROCHIRURGICA, 1992, 118 (3-4) :159-161
[2]   Computer-assisted paranasal sinus surgery using an optoelectronic digitizer system [J].
Hauser, R ;
Westermann, B ;
Reinhardt, H ;
Probst, R .
LARYNGO-RHINO-OTOLOGIE, 1996, 75 (04) :199-207
[3]  
KAVANAGH KT, 1994, LARYNGOSCOPE, V104, P283
[4]   COMPUTER-ASSISTED LOCALIZER FOR PLANNING OF SURGERY AND INTRAOPERATIVE ORIENTATION [J].
LABORDE, G ;
GILSBACH, J ;
HARDERS, A ;
KLIMEK, L ;
MOESGES, R ;
KRYBUS, W .
ACTA NEUROCHIRURGICA, 1992, 119 (1-4) :166-170
[5]  
NITSCHE N, 1993, OTORHINOLARYNGOL NOV, V3, P173
[6]  
NITSCHE N, 1993, OTORHINOLARYNGOL NOV, V3, P57
[7]  
SCHLONDORFF G, 1989, HNO, V37, P187
[8]   A FRAMELESS STEREOTAXIC APPROACH TO NEUROSURGICAL PLANNING BASED ON RETROSPECTIVE PATIENT-IMAGE REGISTRATION [J].
TAN, KK ;
GRZESZCZUK, R ;
LEVIN, DN ;
PELIZZARI, CA ;
CHEN, GTY ;
ERICKSON, RK ;
JOHNSON, D ;
DOHRMANN, GJ .
JOURNAL OF NEUROSURGERY, 1993, 79 (02) :296-303
[9]   Non-invasive 3-D patient registration for image-guided skull base surgery [J].
Westermann, B ;
Hauser, R .
COMPUTERS & GRAPHICS-UK, 1996, 20 (06) :793-799
[10]   OPTICALLY-NAVIGABLE OPERATING MICROSCOPE FOR IMAGE-GUIDED SURGERY [J].
WESTERMANN, B ;
TRIPPEL, M ;
REINHARDT, H .
MINIMALLY INVASIVE NEUROSURGERY, 1995, 38 (03) :112-116