Factors influencina defects of rotary nickel-titanium endodontic instruments after clinical use

被引:204
作者
Parashos, P
Gordon, I
Messer, HH
机构
[1] Univ Melbourne, Sch Dent Sci, Melbourne, Vic 3000, Australia
[2] Univ Melbourne, Stat Consulting Ctr, Melbourne, Vic 3000, Australia
基金
英国医学研究理事会;
关键词
D O I
10.1097/01.DON.0000129963.42882.C9
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
This study examined used, discarded rotary nickel-titanium instruments obtained from 14 endodontists in four countries, and identified factors that may influence defects produced during clinical use. A total of 7,159 instruments were examined for the presence of defects. Unwinding occurred in 12% of instruments and fractures in 5% (1.5% torsional, 3.5% flexural). The defect rates varied Significantly among endodontists. Instrument design factors also influenced defect rate, but to a lesser extent. The mean number of uses of instruments with and without defects was 3.3 +/- 1.8 (range: 1-10), and 4.5 +/- 2.0 (range: 1-16), respectively. The most important influence on defect rates was the operator, which may be related to clinical skill or a conscious decision to use instruments a specified number of times or until defects were evident.
引用
收藏
页码:722 / 725
页数:4
相关论文
共 19 条
[1]   Incidence of rotary ProFile instrument fracture and the potential for bypassing in vivo [J].
Al-Fouzan, KS .
INTERNATIONAL ENDODONTIC JOURNAL, 2003, 36 (12) :864-867
[2]   Scanning electron microscope observations of new and used nickel-titanium rotary files [J].
Alapati, SB ;
Brantley, WA ;
Svec, TA ;
Powers, JM ;
Mitchell, JC .
JOURNAL OF ENDODONTICS, 2003, 29 (10) :667-669
[3]   Evaluation of single-use rotary nickel-titanium instruments [J].
Arens, FC ;
Hoen, MM ;
Steiman, HR ;
Dietz, GC .
JOURNAL OF ENDODONTICS, 2003, 29 (10) :664-666
[4]   Effect of rotational speed on nickel-titanium file distortion [J].
Gabel, WP ;
Hoen, M ;
Steiman, HR ;
Pink, FE ;
Dietz, R .
JOURNAL OF ENDODONTICS, 1999, 25 (11) :752-754
[5]   Cyclic fatigue of Profile rotary instruments after prolonged clinical use [J].
Gambarini, G .
INTERNATIONAL ENDODONTIC JOURNAL, 2001, 34 (05) :386-389
[6]   Fatigue and mechanical properties of nickel-titanium endodontic instruments [J].
Kuhn, G ;
Jordan, L .
JOURNAL OF ENDODONTICS, 2002, 28 (10) :716-720
[7]   Factors influencing the fracture of nickel-titanium rotary instruments [J].
Martín, B ;
Zelada, G ;
Varela, P ;
Bahillo, JG ;
Magán, F ;
Ahn, S ;
Rodríguez, C .
INTERNATIONAL ENDODONTIC JOURNAL, 2003, 36 (04) :262-266
[8]   ProTaper rotary root canal preparation:: assessment of torque and force in relation to canal anatomy [J].
Peters, OA ;
Peters, CI ;
Schönenberger, K ;
Barbakow, F .
INTERNATIONAL ENDODONTIC JOURNAL, 2003, 36 (02) :93-99
[9]   Dynamic torque and apical forces of ProFile .04 rotary instruments during preparation of curved canals [J].
Peters, OA ;
Barbakow, F .
INTERNATIONAL ENDODONTIC JOURNAL, 2002, 35 (04) :379-389
[10]  
Regan JD, 2000, ENDOD DENT TRAUMATOL, V16, P101