Investigation of photon beam output factors for conformal radiation therapy -: Monte Carlo simulations and measurements

被引:77
作者
Haryanto, F
Fippel, M
Laub, W
Dohm, O
Nüsslin, F
机构
[1] Univ Tubingen, Radioonkol Klin, Abt Med Phys, D-72076 Tubingen, Germany
[2] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI 48073 USA
关键词
D O I
10.1088/0031-9155/47/11/401
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to investigate beam output factors (OFs) for conformal radiation therapy and to compare the OFs measured with different detectors with those simulated with Monte Carlo methods. Four different detectors (diode, diamond, pinpoint and ionization chamber) were used to measure photon beam OFs in a water phantom at a depth of 10 cm with a source-surface distance (SSD) of 100 cm. Square fields with widths ranging from 1 cm to 15 cm were observed; the OF for the different field sizes was normalized to that measured at a 5 cm x 5 cm field size at a depth of 10 cm. The BEAM/EGS4 program was used to simulate the exact geometry of a 6 MV photon beam generated by the linear accelerator, and the DOSXYZ-code was implemented to calculate the OFs for all field sizes. Two resolutions (0.1 cm and 0.5 cm voxel size) were chosen here. In addition, to model the detector four kinds of material, water, air, graphite or silicon, were placed in the corresponding voxels. Profiles and depth dose distributions resulting from the simulation show good agreement with the measurements. Deviations of less than 2% can be observed. The OF measured with different detectors in water vary by more than 35% for 1 cm x 1 cm fields. This result can also be found for the simulated OF with different voxel sizes and materials. For field sizes of at least 2 cm x 2 cm the deviations between all measurements and simulations are below 3%. This demonstrates that very small fields have a bad effect on dosimetric accuracy and precision. Finally, Monte Carlo methods can be significant in determining the OF for small fields.
引用
收藏
页码:N133 / N143
页数:11
相关论文
共 24 条
[1]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[2]   A virtual source model for Monte Carlo modeling of arbitrary intensity distributions [J].
Chetty, I ;
DeMarco, JJ ;
Solberg, TD .
MEDICAL PHYSICS, 2000, 27 (01) :166-172
[3]   Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beams [J].
De Vlamynck, K ;
Palmans, H ;
Verhaegen, F ;
De Wagter, C ;
De Neve, W ;
Thierens, H .
MEDICAL PHYSICS, 1999, 26 (09) :1874-1882
[4]   Simple beam models for Monte Carlo photon beam dose calculations in radiotherapy [J].
Fix, MK ;
Keller, H ;
Rüegsegger, P ;
Born, EJ .
MEDICAL PHYSICS, 2000, 27 (12) :2739-2747
[5]   A comparison of dosimetry techniques in stereotactic radiosurgery [J].
Heydarian, M ;
Hoban, PW ;
Beddoe, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) :93-110
[6]   DOSE-RATE DEPENDENCE OF A PTW DIAMOND DETECTOR IN THE DOSIMETRY OF A 6 MV PHOTON-BEAM [J].
HOBAN, PW ;
HEYDARIAN, M ;
BECKHAM, WA ;
BEDDOE, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (08) :1219-1229
[7]   Monitor chamber backscatter for intensity modulated radiation therapy using multileaf collimators [J].
Hounsell, AR .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (02) :445-454
[8]   Monte Carlo calculations of electron beam output factors for a medical linear accelerator [J].
Kapur, A ;
Ma, CM ;
Mok, EC ;
Findley, DO ;
Boyer, AL .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (12) :3479-3494
[9]   Energy and dose rate dependence of a diamond detector in the dosimetry of 4-25 MV photon beams [J].
Laub, WU ;
Kaulich, TW ;
Nusslin, F .
MEDICAL PHYSICS, 1997, 24 (04) :535-536
[10]  
MA CM, 1996, DOSXYZ USERS MANUAL