Immuno-strip biosensor system to detect enrofloxacin residues

被引:28
作者
Kim, Young-Kee [1 ,2 ,3 ]
Kim, Hajin [1 ,2 ]
机构
[1] Hankyong Natl Univ, Dept Chem Engn, Anseong 456749, Kyonggi Do, South Korea
[2] Hankyong Natl Univ, Program Integrated Biotechnol, Anseong 456749, Kyonggi Do, South Korea
[3] Hankyong Natl Univ, Inst Chem Technol, Anseong 456749, Kyonggi Do, South Korea
关键词
Immunosensor; ELISA; Immuno-chromatography assay; Enrofloxacin; Colorimetric; LIQUID-CHROMATOGRAPHY; MASS-SPECTROMETRY; FLUOROQUINOLONE ANTIBIOTICS; FLUORESCENCE DETECTION; CHICKEN MUSCLE; QUINOLONES; PRODUCTS; CLEANUP; KIDNEY; ELISA;
D O I
10.1016/j.jiec.2008.10.007
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The immuno-strip biosensor was developed by combination of immuno-chromatography assay and sandwich enzyme linked immunosorbent assay (ELISA) techniques to detect enrofloxacin, a widely used broad-spectrum antibiotic. A proposed analytical system, consisted of immuno-reaction step and enzyme reaction step, has several advantages such as a very cheap manufacturing cost, a short analysis time (approximately 20 min), an easy-to-use, a high sensitivity, and a qualitatively confirmation by bare eyes. Enzyme reaction step in a proposed analytical tool could enhance the signal of immunosensor, resulting in the high sensitivity. A proposed system was successfully applied to detect enrofloxacin in a range of 100-10000 ppb. The lower detection limit was 100 ppb, which is the maximum residue limit of enrofloxacin in animal product. (C) 2009 The Korean Society of Industrial and Engineering Chemistry. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:229 / 232
页数:4
相关论文
共 24 条
[1]   Liquid chromatographic determination of enrofloxacin in nasal secretions and plasma of healthy pigs using restricted access material for on-line sample clean-up [J].
Bimazubute, M. A. ;
Rozet, E. ;
Dizier, I. ;
Gustin, P. ;
Hubert, Ph. ;
Crommen, J. ;
Chiap, P. .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1189 (1-2) :456-466
[2]   Antibodies to the quinolones and fluoroquinolones for the development of generic and specific immunoassays for detection of these residues in animal products [J].
Bucknall, S ;
Silverlight, J ;
Coldham, N ;
Thorne, L ;
Jackman, R .
FOOD ADDITIVES AND CONTAMINANTS, 2003, 20 (03) :221-228
[3]   Surface plasmon resonance biosensor for enrofloxacin based on deoxyribonucleic acid [J].
Cao, Limin ;
Lin, Hong ;
Mirsky, Vladimir M. .
ANALYTICA CHIMICA ACTA, 2007, 589 (01) :1-5
[4]  
Choi JW, 2006, J MICROBIOL BIOTECHN, V16, P1229
[5]  
*EEC, 1996, J EUR COMMUNITIES L, V125, P10
[6]  
GARLUCCI G, 1998, J CHROMATOGR, V812, P343
[7]  
GILES JS, 1994, NEW MICROBIOL, V19, P155
[8]   Rapid multi-residue screening of antibiotics in muscle and kidney by liquid chromatography-electrospray ionization-tandem mass spectrometry [J].
Granelli, K. ;
Branzell, C. .
ANALYTICA CHIMICA ACTA, 2007, 586 (1-2) :289-295
[9]   Biosensor immunoassay for flumequine in broiler serum and muscle [J].
Haasnoot, Willem ;
Gercek, Husniye ;
Cazemier, Geert ;
Nielen, Michel W. F. .
ANALYTICA CHIMICA ACTA, 2007, 586 (1-2) :312-318
[10]   Determination of residues of quinolones in pig muscle comparative study of classical and microwave extraction techniques [J].
Hermo, MP ;
Barrón, D ;
Barbosa, J .
ANALYTICA CHIMICA ACTA, 2005, 539 (1-2) :77-82