Multicenter evaluation of implantable cardioverter defibrillator testing after implant: The Post Implant Testing Study (PITS)

被引:14
作者
Brodsky, CM
Chang, F
Vlay, SC [1 ]
机构
[1] SUNY Stony Brook, Hlth Sci Ctr T17 020, Div Cardiol, Dept Med, Stony Brook, NY 11794 USA
[2] Columbia Univ, New York, NY USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1999年 / 22卷 / 12期
关键词
defibrillators; implantable; arrhythmia; ventricular fibrillation; electrophysiology; death; sudden; cardiac; cost-benefit analysis; electric countershock;
D O I
10.1111/j.1540-8159.1999.tb00409.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To reassess the function of the implantable cardioverter defibrillator (ICD) many electrophysiology centers perform a second test after the initial test at implant. A prospective multicenter study evaluated the necessity and yield of routine postimplant defibrillator testing. The results of 843 postimplant defibrillator tests were collected from 31 centers. The 764 routine tests in which ventricular fibrillation was successfully induced were analyzed. Variables examined included patient age, presenting arrhythmia, underlying heart disease, left ventricular ejection fraction, defibrillator age, make and model of ICD, electrode system, defibrillation threshold, polarity, and waveform. The overall failure rate was 3.1% (24/764). Units tested later than 365 days after implant tended to have a higher failure rate than those tested within the first month or the next eleven months (6.5%, 3.0%, 2.3%, respectively, P = 0.374). The failure rate was higher in patients with left ventricular ejection fraction <40% than those with higher ejection fractions (3.8% vs 2.0%, P = 0.167). These trends did not reach statistical significance. No other baseline characteristic was associated with higher failure rates. Routine testing of ICDs reveals an overall failure rate of 3.1%. While the rate was low, defibrillator failure places the patient at high risk for sudden cardiac death. As any failure in this population is associated with a high risk of sudden cardiac death, routine defibrillator testing may be justified.
引用
收藏
页码:1769 / 1776
页数:8
相关论文
共 22 条
[1]   Do we need pre-discharge testing in patients who undergo cardioverter-defibrillator implantation? [J].
Achtelik, M ;
Voigt, A ;
Obergassel, L ;
Kattenbeck, K ;
Pfitzner, P ;
Weismüller, P ;
Trappe, HJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (02) :334A-335A
[2]   THE RELATIONSHIPS AMONG VENTRICULAR ARRHYTHMIAS, LEFT-VENTRICULAR DYSFUNCTION, AND MORTALITY IN THE 2 YEARS AFTER MYOCARDIAL-INFARCTION [J].
BIGGER, JT ;
FLEISS, JL ;
KLEIGER, R ;
MILLER, JP ;
ROLNITZKY, LM .
CIRCULATION, 1984, 69 (02) :250-258
[3]  
BRUNN J, 1996, PACE, V19, P605
[4]   EFFECT OF SOTALOL ON VENTRICULAR-FIBRILLATION AND DEFIBRILLATION IN HUMANS [J].
DORIAN, P ;
NEWMAN, D .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 72 (04) :A72-A79
[5]   Utility of predischarge and one-month transvenous implantable defibrillator tests [J].
Goldberger, JJ ;
Horvath, G ;
Inbar, S ;
Kadish, AH .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (06) :822-&
[6]   Value of radiography in diagnosing complications of cardioverter defibrillators implanted without thoracotomy in 437 patients [J].
Gupta, A ;
Zegel, HG ;
Dravid, VS ;
Nierenberg, SJ ;
Freiman, DB .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1997, 168 (01) :105-108
[7]   ICD restudy: Results and potential benefit from routine predischarge and 2-month evaluation [J].
Higgins, SL ;
Rich, DH ;
Haygood, JR ;
Barone, J ;
Greer, SL ;
Meyer, DB .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (02) :410-417
[8]   EFFECTS OF CHRONIC AMIODARONE THERAPY ON DEFIBRILLATION THRESHOLD [J].
JUNG, W ;
MANZ, M ;
PIZZULLI, L ;
PFEIFFER, D ;
LUDERITZ, B .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (11) :1023-1027
[9]   INFLUENCE OF LEFT-VENTRICULAR FUNCTION ON SURVIVAL AND MODE OF DEATH AFTER IMPLANTABLE DEFIBRILLATOR THERAPY (CLEVELAND CLINIC FOUNDATION AND MONTEFIORE MEDICAL-CENTER EXPERIENCE) [J].
KIM, SG ;
MALONEY, JD ;
PINSKI, SL ;
CHOUE, CW ;
FERRICK, KJ ;
ROTH, JA ;
GROSS, J ;
BRODMAN, R ;
FURMAN, S ;
FISHER, JD .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 72 (17) :1263-1267
[10]   Chronic rise in monophasic defibrillation thresholds with a transvenous lead system [J].
Kirk, MM ;
Shorofsky, SR ;
Khalighi, K ;
Kavesh, NG ;
Peters, RW ;
Gold, MR .
AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (04) :502-&