Optical measurement of nebulizer sprays: A quantitative comparison of diffraction, phase Doppler interferometry, and time of flight techniques

被引:38
作者
Corcoran, TE [1 ]
Hitron, R [1 ]
Humphrey, W [1 ]
Chigier, N [1 ]
机构
[1] Carnegie Mellon Univ, Spray Syst Technol Ctr, Pittsburgh, PA 15213 USA
关键词
D O I
10.1016/S0021-8502(99)00030-0
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
Five spray analysis instruments are compared by measurement of identical medical nebulizer sprays. The instruments tested include a phase Doppler interferometer (Aerometrics), a time-of-flight instrument (Amherst Process Instruments) and three diffraction instruments (all by Malvern). Two different jet nebulizers were used for testing. Each was tested with both water and propylene glycol. Volume distributions, mass median aerodynamic diameters (MMAD), and ninety-percent volume aerodynamic diameter measurements (d.(90)) are compared. MMAD results from the Aerosizer were consistently higher than those from the diffraction instruments. MMAD and d.(90) results From the Phase Doppler system agreed well with the diffraction instruments for water sprays, but varied from the other instruments for propylene glycol sprays. This occurred because the Phase Doppler Interferometer reported larger droplets present in the propylene glycol sprays that in some cases significantly affected the otherwise smaller sized volume distributions. These droplets were not apparent in the results from the other instruments bur were noted during visualization studies. Volume distribution results demonstrate that the Malvern 2600 is inappropriate for use with sprays in the examined size range. The other volume distributions were similar with the exception of the Aerosizer, which reported higher, more normally distributed volume distributions compared to the other instruments. (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:35 / 50
页数:16
相关论文
共 30 条
[1]   PHASE DOPPLER SPRAY ANALYZER FOR SIMULTANEOUS MEASUREMENTS OF DROP SIZE AND VELOCITY DISTRIBUTIONS [J].
BACHALO, WD ;
HOUSER, MJ .
OPTICAL ENGINEERING, 1984, 23 (05) :583-590
[2]   CALIBRATION AND PERFORMANCE OF AN API AEROSIZER [J].
CHENG, YS ;
BARR, EB ;
MARSHALL, IA ;
MITCHELL, JP .
JOURNAL OF AEROSOL SCIENCE, 1993, 24 (04) :501-514
[3]   DUAL-BEAM, LIGHT-SCATTERING INTERFEROMETRY FOR SIMULTANEOUS MEASUREMENTS OF DROPLET-SIZE AND VELOCITY DISTRIBUTIONS OF AEROSOLS FROM COMMONLY USED NEBULIZERS [J].
CLIFFORD, RH ;
ISHII, I ;
MONTASER, A ;
MEYER, GA .
ANALYTICAL CHEMISTRY, 1990, 62 (04) :390-394
[4]   Aerosol delivery of liposome-encapsulated ciprofloxacin: Aerosol characterization and efficacy against Francisella tularensis infection in mice [J].
Conley, J ;
Yang, HM ;
Wilson, T ;
Blasetti, K ;
DiNinno, V ;
Schnell, G ;
Wong, JP .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1997, 41 (06) :1288-1292
[5]   DROP-SIZE MEASUREMENT TECHNIQUES FOR SPRAYS - COMPARISON OF MALVERN LASER-DIFFRACTION AND AEROMETRICS PHASE DOPPLER [J].
DODGE, LG ;
RHODES, DJ ;
REITZ, RD .
APPLIED OPTICS, 1987, 26 (11) :2144-2154
[6]   An experimental investigation of the spray issued from a pMDI using laser diagnostic techniques [J].
Dunbar, CA ;
Watkins, AP ;
Miller, JF .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1997, 10 (04) :351-368
[7]   DISTORTION OF CASCADE IMPACTOR MEASURED SIZE DISTRIBUTION DUE TO BOUNCE AND BLOW-OFF [J].
ESMEN, NA ;
LEE, TC .
AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1980, 41 (06) :410-419
[8]   THE INFLUENCE OF SAMPLING CHAMBER DIMENSIONS ON AEROSOL-PARTICLE SIZE MEASUREMENT BY CASCADE IMPACTOR AND TWIN IMPINGER [J].
FULTS, K ;
CYR, TD ;
HICKEY, AJ .
JOURNAL OF PHARMACY AND PHARMACOLOGY, 1991, 43 (10) :726-728
[9]   LUNG DEPOSITION FROM 4 NEBULIZERS [J].
HARDY, JG ;
NEWMAN, SP ;
KNOCH, M .
RESPIRATORY MEDICINE, 1993, 87 (06) :461-465
[10]  
Hesketh H. E., 1986, FINE PARTICLES GASEO