Computed radiography versus screen-film radiography: Detection of pulmonary edema in a rabbit model that simulates neonatal pulmonary infiltrates

被引:13
作者
Don, S [1 ]
Hildebolt, CF [1 ]
Sharp, TL [1 ]
Shackelford, GD [1 ]
Lau, DM [1 ]
Herman, TE [1 ]
McAlister, WH [1 ]
机构
[1] Washington Univ, Sch Med, St Louis Childrens Hosp, Mallinckrodt Inst Technol, St Louis, MO 63110 USA
基金
欧洲研究理事会; 奥地利科学基金会; 欧盟地平线“2020”;
关键词
computed tomography (CT); comparative studies; lung; fluid; radiography; in infants and children; storage phosphor;
D O I
10.1148/radiology.213.2.r99nv15455
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To determine if computed radiography is equivalent to screen-film radiography in depicting pulmonary edema and to determine if radiation exposure can be reduced with computed radiography while maintaining equivalent diagnostic accuracy for pulmonary edema. MATERIALS AND METHODS: Oleic acid was intravenously injected into three rabbits at each of four doses: 0, 0.02, 0.04, and 0.06 mL/kg. Two hours later, chest computed radiographs and screen-film radiographs were obtained at 60 kVp and 1.1 mAs. Additional computed radiographs were obtained after reducing milliampere seconds or by reducing milliampere seconds and increasing the kilovolt peak, which reduced bone marrow exposure by up to 20%. The presence of pulmonary opacities, "truth," was established by the wet-dry weight ratio and by chest computed tomography (CT). The radiographs were masked and randomized. Four observers rated the images for the presence of parenchymal opacities with a dichotomous score and judged the quality of the radiographs on a scale from 1 (worst) to 6 (best). Cochran Q tests and McNemar tests were used to analyze the differences in paired comparisons. Image quality was evaluated with logistic regression analysis. RESULTS: There was no significant difference between truth and observer ability to detect opacity for either modality or for any exposure (P > .05). There was no significant difference between computed radiography and screen-film radiography for image quality (P > .05). CONCLUSION: Computed radiography is equivalent to screen-film radiography in the detection of pulmonary edema. Radiation exposure reduction of 20% can be achieved without affecting pulmonary edema detection or image quality.
引用
收藏
页码:455 / 460
页数:6
相关论文
共 18 条
[1]  
BOGUCKI T, 1996, TECHNICAL SCI MONOGR, V6
[2]   PHOSPHOR PLATE COMPUTED RADIOGRAPHY - RESPONSE TO VARIATION IN MAS AT FIXED KVP IN AN ANIMAL-MODEL - POTENTIAL ROLE IN NEONATAL IMAGING [J].
BRODERICK, NJ ;
LONG, B ;
DREESEN, RG ;
COHEN, MD ;
CORY, DA ;
KATZ, BP ;
KALASINSKI, LA .
CLINICAL RADIOLOGY, 1993, 47 (01) :39-45
[3]   COMPARISON OF COMPUTERIZED DIGITAL AND FILM-SCREEN RADIOGRAPHY - RESPONSE TO VARIATION IN IMAGING KVP [J].
BRODERICK, NJ ;
LONG, B ;
DREESEN, RG ;
COHEN, MD ;
CORY, DA ;
KATZ, BP ;
KALASINSKI, LA .
PEDIATRIC RADIOLOGY, 1992, 22 (05) :346-349
[4]   DIGITAL IMAGING WITH A PHOTOSTIMULABLE PHOSPHOR IN THE CHEST OF NEWBORNS [J].
COHEN, MD ;
KATZ, BP ;
KALASINSKI, LA ;
WHITE, SJ ;
SMITH, JA ;
LONG, B .
RADIOLOGY, 1991, 181 (03) :829-832
[5]  
CURRY TS, 1984, CHRISTENSENS INTRO P, P401
[6]  
CURRY TS, 1984, CHRISTENSENS INTRO P, P161
[7]  
DERKS CM, 1977, AM J PATHOL, V87, P143
[8]   VOLUME DETECTION THRESHOLD - QUANTITATIVE COMPARISON OF COMPUTED RADIOGRAPHY AND SCREEN-FILM RADIOGRAPHY IN DETECTION OF PNEUMOTHORACES IN AN ANIMAL-MODEL THAT SIMULATES THE NEONATE [J].
DON, S ;
COHEN, MD ;
KRUGER, RA ;
WINKLER, TA ;
KATZ, BP ;
LI, W ;
DREESEN, RG ;
KENNAN, N ;
TARVER, R ;
KLATTE, EC .
RADIOLOGY, 1995, 194 (03) :727-730
[9]   RABBIT PNEUMOTHORAX - FEASIBILITY OF AN ANIMAL-MODEL TO SIMULATE NEONATAL PNEUMOTHORAX [J].
DON, S ;
COHEN, MD ;
KRUGER, RA ;
WINKLER, TA ;
KATZ, BP ;
LI, W ;
DREESEN, G ;
KENNAN, N ;
TARVER, R ;
KLATTE, EC .
INVESTIGATIVE RADIOLOGY, 1994, 29 (05) :558-563
[10]  
Hillman B J, 1989, J Digit Imaging, V2, P220