Detection of a Meal Using Continuous Glucose Monitoring implications for an artificial β-cell

被引:87
作者
Dassau, Eyal [1 ]
Bequette, B. Wayne [2 ]
Buckingham, Bruce A. [3 ]
Doyle, Francis J., III [1 ]
机构
[1] Univ Calif Santa Barbara, Dept Chem Engn, Santa Barbara, CA 93106 USA
[2] Rensselaer Polytech Inst, Dept Chem & Biol Engn, Troy, NY USA
[3] Stanford Med Ctr, Stanford, CA USA
基金
美国国家卫生研究院;
关键词
SYSTEM; MODEL; CHILDREN; TIME;
D O I
10.2337/dc07-1293
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVE - The purpose of this Study -,vas to introduce a novel meal detection algorithm (MDA) to be used as part of an artificial P-cell that uses a continuous glucose monitor (CGM). RESEARCH DESIGN AND METHODS - We developed our MDA on a dataset of 26 meal events using records from 19 children aged 1-6 years who Used the MiniMed CGMS Gold. We then applied this algorithm to CGM records from a DirecNet pilot study of the FreeStyle Navigator continuous glucose sensor. During a research center admission, breakfast insulin was withheld for 1 h, and discrete glucose levels were obtained every 10 min after the meal. RESULTS - Based on the Navigator readings, the MDA detected a meal at a mean time of 30 min from the onset of eating, at which time the mean serum glucose was 21 mg/dl above baseline (range 2-36 mg/dl), and >90% of meals were detected before the glucose had risen 40 mg/dl from baseline. CONCLUSIONS - The MDA will enable automated insulin dosing in response to meals, facilitating the development of an artificial pancreas.
引用
收藏
页码:295 / 300
页数:6
相关论文
共 21 条
  • [1] Bequette B Wayne, 2005, Diabetes Technol Ther, V7, P28, DOI 10.1089/dia.2005.7.28
  • [2] Bequette BW, 2004, P AMER CONTR CONF, P958
  • [3] A system model of oral glucose absorption: Validation on gold standard data
    Dalla Man, Chiara
    Camilleri, Michael
    Cobelli, Claudio
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (12) : 2472 - 2478
  • [4] Dassau E, 2007, DIABETES, V56, pA23
  • [5] The medtronic MiniMed gold continuous glucose monitoring system: An effective means to discover hypo- and Hyperglycemia in children under 7 years of age
    Gandrud, Laura M.
    Xing, Dongyuan
    Kollman, Craig
    Block, Jen M.
    Kunselman, Betsy
    Wilson, Darrell M.
    Buckingham, Bruce A.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2007, 9 (04) : 307 - 316
  • [6] A new approach to diabetic control: Fuzzy logic and insulin pump technology
    Grant, Paul
    [J]. MEDICAL ENGINEERING & PHYSICS, 2007, 29 (07) : 824 - 827
  • [7] Continuous glucose monitoring and closed-loop systems
    Hovorka, R
    [J]. DIABETIC MEDICINE, 2006, 23 (01) : 1 - 12
  • [8] Roadmap to the artificial pancreas
    Hovorka, Roman
    Wilinska, Malgorzata E.
    Chassin, Ludovic J.
    Dunger, David B.
    [J]. DIABETES RESEARCH AND CLINICAL PRACTICE, 2006, 74 : S178 - S182
  • [9] Knobbe Edward J, 2005, Diabetes Technol Ther, V7, P15, DOI 10.1089/dia.2005.7.15
  • [10] KUUREKINSEY CR, 2006, P AM CONTR C MINN MN, P2957