Characterization of a mammographic system based on single photon counting pixel arrays coupled to GaAs x-ray detectors

被引:12
作者
Amendolia, S. R. [2 ,3 ]
Bisogni, M. G. [1 ,2 ]
Delogu, P. [1 ,2 ]
Fantacci, M. E. [1 ,2 ]
Paternoster, G.
Rosso, V. [1 ,2 ]
Stefanini, A. [1 ,2 ]
机构
[1] Univ Pisa, Dip Fis E Fermi, I-56127 Pisa, Italy
[2] INFN, Sez Pisa, I-56127 Pisa, Italy
[3] Univ Sassari, Str Dip Matemat & Fis, I-07100 Sassari, Italy
关键词
biomedical equipment; diagnostic radiography; image resolution; mammography; optical transfer function; photon counting; position sensitive particle detectors; semiconductor counters; X-ray detection; READOUT CHIP; PERFORMANCE; SIMULATION; MEDIPIX2; WORKING;
D O I
10.1118/1.3097284
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
The authors report on the imaging capabilities of a mammographic system demonstrator based on GaAs pixel detectors operating in single photon counting (SPC) mode. The system imaging performances have been assessed by means of the transfer functions: The modulation transfer function (MTF), the normalized noise power spectrum, and the detective quantum efficiency (DQE) have been measured following the guidelines of the IEC 62220-1-2 protocol. The transfer function analysis has shown the high spatial resolution capabilities of the GaAs detectors. The MTF calculated at the Nyquist frequency (2.94 cycles/mm) is indeed 60%. The DQE, measured with a standard mammographic beam setup (Mo/Mo, 28 kVp, with 4 mm Al added filter) and calculated at zero frequency, is 46%. Aiming to further improve the system's image quality, the authors investigate the DQE limiting factors and show that they are mainly related to system engineering. For example, the authors show that optimization of the image equalization procedure increases the DQE(0) up to 74%, which is better than the DQE(0) of most clinical mammographic systems. The authors show how the high detection efficiency of GaAs detectors and the noise discrimination associated with the SPC technology allow optimizing the image quality in mammography. In conclusion, the authors propose technological solutions to exploit to the utmost the potentiality of GaAs detectors coupled to SPC electronics.
引用
收藏
页码:1330 / 1339
页数:10
相关论文
共 33 条
[1]
Experimental study of Compton scattering reduction in digital mammographic imaging [J].
Amendolia, S.R. ;
Bisogni, M.G. ;
Delogu, P. ;
Fantacci, M.E. ;
Novelli, M. ;
Oliva, P. ;
Quattrocchi, M. ;
Rosso, V. ;
Stefanini, A. ;
Zucca, Sergio .
IEEE Transactions on Nuclear Science, 2002, 49 I (05) :2361-2365
[2]
Low contrast imaging with a GaAs pixel digital detector [J].
Amendolia, SR ;
Bisogni, MG ;
Bottigli, U ;
Ciocci, MA ;
Delogu, P ;
Dipasquale, G ;
Fantacci, ME ;
Giannelli, M ;
Maestro, P ;
Marzulli, VM ;
Pernigotti, E ;
Rosso, V ;
Stefanini, A ;
Stumbo, S .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (04) :1478-1482
[3]
A GaAs pixel detectors-based digital mammographic system: Performances and imaging tests results [J].
Annovazzi, A. ;
Amendolia, S. R. ;
Bigongiari, A. ;
Bisogni, M. G. ;
Catarsi, F. ;
Cesqui, F. ;
Cetronio, A. ;
Colombo, F. ;
Delogu, P. ;
Fantacci, M. E. ;
Gilberti, A. ;
Lanzieri, C. ;
Lavagna, S. ;
Novelli, M. ;
Passuello, G. ;
Paternoster, G. ;
Pieracci, M. ;
Poletti, M. ;
Quattrocchi, M. ;
Rosso, V. ;
Stefanini, A. ;
Testa, A. ;
Venturelli, L. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 576 (01) :154-159
[4]
Physical characterization of a scanning photon counting digital mammography system based on Si-strip detectors [J].
Aslund, Magnus ;
Cederstrom, Bjorn ;
Lundqvist, Mats ;
Danielsson, Mats .
MEDICAL PHYSICS, 2007, 34 (06) :1918-1925
[5]
The medipix3 prototype, a pixel readout chip working in single photon counting mode with improved spectrometric performance [J].
Ballabriga, R. ;
Campbell, M. ;
Heijne, E. H. M. ;
Llopart, X. ;
Tlustos, L. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (05) :1824-1829
[7]
Performance of a 4096 pixel photon counting chip [J].
Bisogni, MG ;
Campbell, M ;
Conti, M ;
Delogu, P ;
Fantacci, ME ;
Heijne, EHM ;
Maestro, P ;
Magistrati, G ;
Marzulli, VM ;
Meddeler, G ;
Mikulec, B ;
Pernigotti, E ;
Rosso, V ;
Schwarz, C ;
Snoeys, W ;
Stumbo, S ;
Watt, J .
EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY IX, 1998, 3445 :298-304
[8]
BISOGNI MG, 2004, P IEEE NUCL SC UNPUB, V7, P4524
[9]
Dear-Mama:: A photon counting X-ray imaging project for medical applications [J].
Blanchot, G. ;
Chmeissani, M. ;
Diaz, A. ;
Diaz, F. ;
Fernandez, J. ;
Garcia, E. ;
Garcia, J. ;
Kainberger, F. ;
Lozano, M. ;
Maiorina, M. ;
Martinez, R. ;
Montagne, J. P. ;
Moreno, I. ;
Pellegrini, G. ;
Puigdengoles, C. ;
Sentis, M. ;
Teres, L. ;
Tortajada, M. ;
Ullan, M. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (01) :136-139
[10]
Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image [J].
Buhr, E ;
Günther-Kohfahl, S ;
Neitzel, U .
MEDICAL PHYSICS, 2003, 30 (09) :2323-2331