Physical characterization of a scanning photon counting digital mammography system based on Si-strip detectors

被引:75
作者
Aslund, Magnus [1 ]
Cederstrom, Bjorn
Lundqvist, Mats
Danielsson, Mats
机构
[1] Royal Inst Technol, Dept Phys, AlbaNova, S-10691 Stockholm, Sweden
[2] Sectra Mamea AB, S-16440 Kista, Sweden
[3] Royal Inst Technol, Dept Phys, S-10691 Stockholm, Sweden
关键词
mammography; detective quantum efficiency; scanning;
D O I
10.1118/1.2731032
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The physical performance of a scanning multislit full field digital mammography system was determined using basic image quality parameters. The system employs a direct detection detector comprised of linear silicon strip sensors in an edge-on geometry connected to photon counting electronics. The pixel size is 50 mu m and the field of view 24 x 26 cm(2). The performance was quantified using the presampled modulation transfer function, the normalized noise power spectrum and the detective quantum efficiency (DQE). Compared to conventional DQE methods, the scanning geometry with its intrinsic scatter rejection poses additional requirements on the measurement setup, which are investigated in this work. The DQE of the photon counting system was found to be independent of the dose level to the detector in the 7.6-206 mu Gy range. The peak DQE was 72% and 73% in the scan and slit direction, respectively, measured with a 28 kV W-0.5 mm Al anodefilter combination with an added 2 mm Al filtration. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1918 / 1925
页数:8
相关论文
共 28 条
[1]   Scatter rejection in multislit digital mammography [J].
Åslund, M ;
Cederström, B ;
Lundqvist, M ;
Danielsson, M .
MEDICAL PHYSICS, 2006, 33 (04) :933-940
[2]  
ASLUND M, 2006, P SPIE, V6142
[3]   Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography [J].
Boone, JM ;
Fewell, TR ;
Jennings, RJ .
MEDICAL PHYSICS, 1997, 24 (12) :1863-1874
[4]   Comparison of x-ray cross sections for diagnostic and therapeutic medical physics [J].
Boone, JM ;
Chavez, AE .
MEDICAL PHYSICS, 1996, 23 (12) :1997-2005
[5]  
BURNS PD, 1987, P SPIE, V767
[6]   Detective quantum efficiency dependence on x-ray energy weighting in mammography [J].
Cahn, RN ;
Cederström, B ;
Danielsson, M ;
Hall, A ;
Lundqvist, M ;
Nygren, D .
MEDICAL PHYSICS, 1999, 26 (12) :2680-2683
[7]   Signal-to-noise optimization of medical imaging systems [J].
Cunningham, IA ;
Shaw, R .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1999, 16 (03) :621-632
[8]   Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol [J].
Dance, DR ;
Skinner, CL ;
Young, KC ;
Beckett, JR ;
Kotre, CJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (11) :3225-3240
[9]   Intercomparison of methods for image quality characterization. II. Noise power spectrum [J].
Dobbins, James T., III ;
Samei, Ehsan ;
T Ranger, Nicole ;
Chen, Ying .
MEDICAL PHYSICS, 2006, 33 (05) :1466-1475
[10]   Performance evaluation of a "dual-side read" dedicated mammography computed radiography system [J].
Fetterly, KA ;
Schueler, BA .
MEDICAL PHYSICS, 2003, 30 (07) :1843-1854