AC voltammetric carbon paste-based enzyme immunosensors

被引:56
作者
Fernández-Sánchez, C [1 ]
González-García, MB [1 ]
Costa-García, A [1 ]
机构
[1] Univ Oviedo, Dept Quim Fis & Analit, Oviedo 33006, Asturias, Spain
关键词
carbon paste electrode; alkaline phosphatase; 3-Indoxyl phosphate; indigo; alternating current voltammetry;
D O I
10.1016/S0956-5663(99)00065-2
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Carbon paste electrodes, previously anodised in a basic media. are the basis for the development of a new voltammetric immunosensor device. Passive adsorption of the appropriate immunochemical reagent was performed onto the electrode surface. Alkaline Phosphatase labelled immunoglobulin was the tracer used in this work, 3-indoxyl phosphate being a very suitable enzymatic substrate for the electrochemical detection of the corresponding affinity reaction. The hydrolysis of this molecule generates indigo dimmer. This product was detected by alternating current voltammetry taking advantage of the adsorptive and inherent electrodic properties that it exhibits. The same electrochemical anodisation was used at the end of one assay to remove the entire protein layer attached to the carbon paste surface, allowing the formation of a new sensing phase and the use of the same support in several consecutive experiments. The methodology was applied to the design of two different immunoassays for the determination of human IgG. Good reproducibility of the electrodic signal and a limit of detection around 10(-10) M were achieved. (C) 2000 Elsevier Science S.A. All rights reserved.
引用
收藏
页码:917 / 924
页数:8
相关论文
共 30 条
[11]   Adsorption of immunoglobulin G on carbon paste electrodes as a basis for the development of immunoelectrochemical devices [J].
FernandezSanchez, C ;
CostaGarcia, A .
BIOSENSORS & BIOELECTRONICS, 1997, 12 (05) :403-413
[12]  
Gonzalez Garcia M. B., 1995, Bioelectrochemistry and Bioenergetics, V38, P389, DOI 10.1016/0302-4598(95)01813-T
[13]   ENZYMATIC ELECTROCATALYSIS AS A STRATEGY FOR ELECTROCHEMICAL DETECTION IN HETEROGENEOUS IMMUNOASSAYS [J].
GYSS, C ;
BOURDILLON, C .
ANALYTICAL CHEMISTRY, 1987, 59 (19) :2350-2355
[14]   AUTOMATIC-APPARATUS FOR HETEROGENEOUS ENZYME IMMUNOASSAYS BASED ON ELECTROCATALYTIC DETECTION OF THE ENZYME AND ELECTROCHEMICAL REGENERATION OF THE SOLID-PHASE [J].
HUET, D ;
BOURDILLON, C .
ANALYTICA CHIMICA ACTA, 1993, 272 (02) :205-212
[15]   DEVELOPMENT OF A POLYPYRROLE-BASED HUMAN SERUM-ALBUMIN SENSOR [J].
JOHN, R ;
SPENCER, M ;
WALLACE, GG ;
SMYTH, MR .
ANALYTICA CHIMICA ACTA, 1991, 249 (02) :381-385
[16]   Fast amperometric immunoassay for hantavirus infection [J].
Krishnan, R ;
Ghindilis, AL ;
Atanasov, P ;
Wilkins, E ;
Montoya, J ;
Koster, FT .
ELECTROANALYSIS, 1996, 8 (12) :1131-1134
[17]   Ferrocenylethyl phosphate: An improved substrate for the detection of alkaline phosphatase by cathodic stripping ion-exchange voltammetry. Application to the electrochemical enzyme affinity assay of avidin [J].
Limoges, B ;
Degrand, C .
ANALYTICAL CHEMISTRY, 1996, 68 (23) :4141-4148
[18]   Immunological activities of IgG antibody on pre-coated Fc receptor surfaces [J].
Lu, B ;
Smyth, MR ;
OKennedy, R .
ANALYTICA CHIMICA ACTA, 1996, 331 (1-2) :97-102
[19]   SMALL-VOLUME VOLTAMMETRIC DETECTION OF 4-AMINOPHENOL WITH INTERDIGITATED ARRAY ELECTRODES AND ITS APPLICATION TO ELECTROCHEMICAL ENZYME-IMMUNOASSAY [J].
NIWA, O ;
XU, Y ;
HALSALL, HB ;
HEINEMAN, WR .
ANALYTICAL CHEMISTRY, 1993, 65 (11) :1559-1563
[20]   Amperometric immunosensing using microperoxidase MP-11 antibody conjugates [J].
Padeste, C ;
Grubelnik, A ;
Tiefenauer, L .
ANALYTICA CHIMICA ACTA, 1998, 374 (2-3) :167-176