Impact of final Stent dimensions on long-term results following sirolimus-eluting stent implantation - Serial intravascular ultrasound analysis from the SIRIUS trial
被引:342
作者:
论文数: 引用数:
h-index:
机构:
Sonoda, S
Morino, Y
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Morino, Y
Ako, J
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Ako, J
Terashima, M
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Terashima, M
Hassan, AHM
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Hassan, AHM
Bonneau, HN
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Bonneau, HN
Leon, MB
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Leon, MB
Moses, JW
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Moses, JW
Yock, PG
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Yock, PG
Honda, Y
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Honda, Y
Kuntz, RE
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Kuntz, RE
Fitzgerald, PJ
论文数: 0引用数: 0
h-index: 0
机构:Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
Fitzgerald, PJ
机构:
[1] Stanford Univ, Ctr Med, Ctr Res Cardiovasc Intervent, Stanford, CA 94305 USA
OBJECTIVES We assessed the predictive value of minimum stent area (MSA) for long-term patency of sirolimus-eluting stents (SES) implantation compared to bare metal stents (BMS). BACKGROUND Although MSA is a consistent predictor of in-stent restenosis, its predictive value in BMS is still limited because of biologic variability in the restenosis process. METHODS From the SIRolImUS (SIRIUS) trial, 122 cases (SES: 72; BMS: 50) with complete serial intravascular ultrasound (IVUS) (baseline and 8-month follow-up) were analyzed. Postprocedure MSA and follow-up minimum lumen area (MLA) were obtained. Based on previous physiologic studies, adequate stent patency at follow-up was defined as MLA >4 mm(2). RESULTS In both groups, a significant positive correlation was observed between baseline MSA and follow-up MLA (SES: p < 0.0001, BMS: p < 0.0001). However, SES showed higher correlation than BMS (0.8 vs. 0.65) with a higher regression coefficient (0.92 vs. 0.59). The sensitivity and specificity curves identified different optimal thresholds of MSA to predict adequate follow-up MLA: 5 mm(2) for SES and 6.5 mm(2) for BMS. The positive predictive values with these cutoff points were 90% and 56%, respectively. CONCLUSIONS In this SIRIUS IVUS substudy, SES reduced both biologic variability and restenosis, resulting in increased predictability of long-term stent patency with postprocedure MSA. In addition, SES had a considerably lower optimal MSA threshold compared to BMS. (C) 2004 by the American College of Cardiology Foundation.