摘要
A method for evaluating the overall uncertainty of alanine EPR transfer dosimetry in the therapeutic dose range is described. The method uses experimental data on EPR signal reproducibility from replicate dosimeters irradiated to low doses (1-5Gy), estimates of Type B uncertainties, and Monte Carlo simulations of heteroscedastic orthogonal linear regression. A Bruker ECS106 spectrometer and Bruker alanine dosimeters have been used for this evaluation. The results demonstrate that alanine dosimetry can be used for transfer dosimetry in that range with the overall uncertainty 1.5-4% (1sigma) depending on the dose. the number of replicate dosimeters. and the duration of the calibration session (the session should not exceed one working day), Published by Elsevier Science Ltd.
引用
收藏
页码:917 / 929
页数:13
相关论文
共 44 条
[1]  
AHLERS FJ, 1991, RADIAT PROT DOSIM, V37, P117
[2]   SOURCES OF UNCERTAINTY IN THERAPY LEVEL ALANINE DOSIMETRY [J].
BARTOLOTTA, A ;
FATTIBENE, P ;
ONORI, S ;
PANTALONI, M ;
PETETTI, E .
APPLIED RADIATION AND ISOTOPES, 1993, 44 (1-2) :13-17
[3]   ESR solid state dosimetry: Behaviour of various amino acids and blend preparation procedures [J].
Bartolotta, A ;
Brai, M ;
Caputo, V ;
De Caro, V ;
Giannola, LI ;
Rap, R ;
Teri, G .
RADIATION PROTECTION DOSIMETRY, 1999, 84 (1-4) :293-296
[4]   ARE THERE 2 REGRESSIONS [J].
BERKSON, J .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1950, 45 (250) :164-180
[5]   Application of high precision EPR dosimetry with teeth for reconstruction of doses to Chernobyl populations [J].
Chumak, V ;
Sholom, S ;
Pasalskaya, L .
RADIATION PROTECTION DOSIMETRY, 1999, 84 (1-4) :515-520
[6]   An approach to the assessment of overall uncertainty of determination of dose using an ESR technique [J].
Chumak, V ;
Pavlenko, J ;
Sholom, S .
APPLIED RADIATION AND ISOTOPES, 1996, 47 (11-12) :1287-1291
[7]   ENERGY RESPONSE OF AGAR-ALANINE FREE-RADICAL DOSIMETRY TO THERAPEUTIC ELECTRON-BEAMS [J].
CIESIELSKI, B ;
REINSTEIN, LE ;
MEEK, AG ;
WIELOPOLSKI, L .
MEDICAL PHYSICS, 1993, 20 (05) :1453-1455
[8]   THE EFFECTS OF BORON ON THE ELECTRON-PARAMAGNETIC-RESONANCE SPECTRA OF ALANINE IRRADIATED WITH THERMAL-NEUTRONS [J].
CIESIELSKI, B ;
WIELOPOLSKI, L .
RADIATION RESEARCH, 1995, 144 (01) :59-63
[9]  
CREASY MA, 1954, J ROY STAT SOC B, V16, P186
[10]  
Cuttone G, 1999, PHYS MEDICA, V15, P121